Transthoracic impedance (TTI) is an important determinant of success in defibrillation. Low TTI increases transmyocardial current and therefore increases the chance of depolarising a critical mass of myocardium. A major component of TTI occurs at the paddle-skin interface and is minimised by pressure applied to the defibrillation paddles. The International Liaison Committee on Resuscitation (ILCOR) 2000 guidelines recommend that 'firm force' should be applied to both paddles, whereas previous European Resuscitation Council (ERC) 1992 guidelines were more precise, recommending that 12 kg of force should to be applied. We assessed whether defibrillator operators are capable of achieving 12 kg paddle force. Fifty advanced life support-trained doctors and nurses attempted to achieve 12 kg paddle force while simulating defibrillation on a resuscitation doll. The median value of the maximum pressures obtainable was 10.1 (max 16.0; min 5.0) kg force. Only 14% could achieve > or =12 kg force on both paddles for defibrillation. Men achieved more force than women (10.7 vs. 8.1 kg force; P<0.01), and there was a correlation between maximum force achieved and operator height (r2=0.27) and dominant hand-grip strength (r2=0.34). The ERC recommendation of 12 kg paddle force is not achievable by the majority of defibrillator operators.
defibrillation, paddle force, transthoracic impedance, guidelines
287-290
Sado, D.M.
0ba2017a-85b2-46d1-b0d8-1e54bfddf827
Deakin, C.D.
560d993b-bbc9-4548-9990-272ed18a011d
Petley, G.W.
4f2da40b-3c7b-4adc-b75c-e24e62bb1cf0
December 2001
Sado, D.M.
0ba2017a-85b2-46d1-b0d8-1e54bfddf827
Deakin, C.D.
560d993b-bbc9-4548-9990-272ed18a011d
Petley, G.W.
4f2da40b-3c7b-4adc-b75c-e24e62bb1cf0
Abstract
Transthoracic impedance (TTI) is an important determinant of success in defibrillation. Low TTI increases transmyocardial current and therefore increases the chance of depolarising a critical mass of myocardium. A major component of TTI occurs at the paddle-skin interface and is minimised by pressure applied to the defibrillation paddles. The International Liaison Committee on Resuscitation (ILCOR) 2000 guidelines recommend that 'firm force' should be applied to both paddles, whereas previous European Resuscitation Council (ERC) 1992 guidelines were more precise, recommending that 12 kg of force should to be applied. We assessed whether defibrillator operators are capable of achieving 12 kg paddle force. Fifty advanced life support-trained doctors and nurses attempted to achieve 12 kg paddle force while simulating defibrillation on a resuscitation doll. The median value of the maximum pressures obtainable was 10.1 (max 16.0; min 5.0) kg force. Only 14% could achieve > or =12 kg force on both paddles for defibrillation. Men achieved more force than women (10.7 vs. 8.1 kg force; P<0.01), and there was a correlation between maximum force achieved and operator height (r2=0.27) and dominant hand-grip strength (r2=0.34). The ERC recommendation of 12 kg paddle force is not achievable by the majority of defibrillator operators.
This record has no associated files available for download.
More information
Published date: December 2001
Keywords:
defibrillation, paddle force, transthoracic impedance, guidelines
Organisations:
Clinical & Experimental Sciences
Identifiers
Local EPrints ID: 383380
URI: http://eprints.soton.ac.uk/id/eprint/383380
ISSN: 0300-9572
PURE UUID: dd82912d-a11e-4142-b06a-5ca91763e206
Catalogue record
Date deposited: 29 Jan 2016 09:55
Last modified: 15 Mar 2024 02:53
Export record
Altmetrics
Contributors
Author:
D.M. Sado
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics