Nondestructive evaluation of bone cement and bone cement/metal interface failure
Nondestructive evaluation of bone cement and bone cement/metal interface failure
To quantify the failure mechanisms related to the loosening of cemented hip joint
replacements, novel techniques, capable of monitoring, nondestructively, the initiation and
progression of failure during in vitro fatigue tests, were employed. Fatigue testing of model
cement and cement-stem test pieces was monitored using acoustic emission (AE) sensors. Once
damage was detected, an ultrasonic imaging system was used to obtain an image of the damage
site and to measure the stiffness of the affected region. This method of examination provided a
detailed insight into the internal crack propagation and delamination patterns. Initial work
was conducted on bulk cement specimens subjected to bending and tension. The second stage
of the work examined a model stem-cement interface under tensile opening loading conditions.
A novel ultrasonic technique was used to measure the bond quality at the cement-metal
interface. Progressive delamination was identified over time, and the AE technique was able to
identify critical areas of delamination before they could be identified conclusively by
ultrasonic imaging. The work has demonstrated the potential of the AE technique as a tool for
the preclinical assessment of total hip replacements.
acoustic emission, ultrasonics, bone cement, interfacial failure
420-429
Browne, M.
6578cc37-7bd6-43b9-ae5c-77ccb7726397
Jeffers, J.R.T.
a4a85d45-09ac-40ce-99d3-0ceb3bb3d434
Saffari, N.
5a5dd869-96f3-4f81-ae25-840c800ca969
February 2010
Browne, M.
6578cc37-7bd6-43b9-ae5c-77ccb7726397
Jeffers, J.R.T.
a4a85d45-09ac-40ce-99d3-0ceb3bb3d434
Saffari, N.
5a5dd869-96f3-4f81-ae25-840c800ca969
Browne, M., Jeffers, J.R.T. and Saffari, N.
(2010)
Nondestructive evaluation of bone cement and bone cement/metal interface failure.
Journal of Biomedical Materials Research Part B: Applied Biomaterials, 92B, .
(doi:10.1002/jbm.b.31530).
Abstract
To quantify the failure mechanisms related to the loosening of cemented hip joint
replacements, novel techniques, capable of monitoring, nondestructively, the initiation and
progression of failure during in vitro fatigue tests, were employed. Fatigue testing of model
cement and cement-stem test pieces was monitored using acoustic emission (AE) sensors. Once
damage was detected, an ultrasonic imaging system was used to obtain an image of the damage
site and to measure the stiffness of the affected region. This method of examination provided a
detailed insight into the internal crack propagation and delamination patterns. Initial work
was conducted on bulk cement specimens subjected to bending and tension. The second stage
of the work examined a model stem-cement interface under tensile opening loading conditions.
A novel ultrasonic technique was used to measure the bond quality at the cement-metal
interface. Progressive delamination was identified over time, and the AE technique was able to
identify critical areas of delamination before they could be identified conclusively by
ultrasonic imaging. The work has demonstrated the potential of the AE technique as a tool for
the preclinical assessment of total hip replacements.
Text
Browne_et_al_AE_interface.pdf
- Version of Record
Restricted to Repository staff only
Request a copy
More information
Published date: February 2010
Keywords:
acoustic emission, ultrasonics, bone cement, interfacial failure
Identifiers
Local EPrints ID: 143229
URI: http://eprints.soton.ac.uk/id/eprint/143229
ISSN: 1552-4973
PURE UUID: c8d07ece-3e8c-481a-b389-eb6b421ab509
Catalogue record
Date deposited: 08 Apr 2010 08:43
Last modified: 14 Mar 2024 02:39
Export record
Altmetrics
Contributors
Author:
J.R.T. Jeffers
Author:
N. Saffari
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.