Effect of geometrical uncertainty on cemented hip implant structural integrity
Effect of geometrical uncertainty on cemented hip implant structural integrity
A large number of parameters such as material properties, geometry, and structural strength are involved in the design and analysis of cemented hip implants. Uncertainties in these parameters have a potential to compromise the structural performance and lifetime of implants. Statistical analyses are well suited to investigating this type of problem as they can estimate the influence of these uncertainties on the incidence of failure. Recent investigations have focused on the effect of uncertainty in cement properties and loading condition on the integrity of the construct. The present study hypothesizes that geometrical uncertainties will play a role in cement mantle failure. Finite element input parameters were simulated as random variables and different modes of failure were investigated using a response surface method (RSM). The magnitude of random von Mises stresses varied up to 8 MPa, compared with a maximum nominal value of 2.38 MPa. Results obtained using RSM are shown to match well with a benchmark direct Monte Carlo simulation method. The resulting probability that the maximum cement stress will exceed the nominal stress is 62%. The load and the bone and prosthesis geometries were found to be the parameters most likely to influence the magnitude of the cement stresses and therefore to contribute most to the probability of failure
054501-[5pp]
Bah, Mamadou T.
b5cd0f47-016f-485c-8293-5f6bf8a7ef1a
Browne, Martin
6578cc37-7bd6-43b9-ae5c-77ccb7726397
May 2009
Bah, Mamadou T.
b5cd0f47-016f-485c-8293-5f6bf8a7ef1a
Browne, Martin
6578cc37-7bd6-43b9-ae5c-77ccb7726397
Bah, Mamadou T. and Browne, Martin
(2009)
Effect of geometrical uncertainty on cemented hip implant structural integrity.
Journal of Biomechanical Engineering, 131 (5), .
(doi:10.1115/1.3078172).
Abstract
A large number of parameters such as material properties, geometry, and structural strength are involved in the design and analysis of cemented hip implants. Uncertainties in these parameters have a potential to compromise the structural performance and lifetime of implants. Statistical analyses are well suited to investigating this type of problem as they can estimate the influence of these uncertainties on the incidence of failure. Recent investigations have focused on the effect of uncertainty in cement properties and loading condition on the integrity of the construct. The present study hypothesizes that geometrical uncertainties will play a role in cement mantle failure. Finite element input parameters were simulated as random variables and different modes of failure were investigated using a response surface method (RSM). The magnitude of random von Mises stresses varied up to 8 MPa, compared with a maximum nominal value of 2.38 MPa. Results obtained using RSM are shown to match well with a benchmark direct Monte Carlo simulation method. The resulting probability that the maximum cement stress will exceed the nominal stress is 62%. The load and the bone and prosthesis geometries were found to be the parameters most likely to influence the magnitude of the cement stresses and therefore to contribute most to the probability of failure
This record has no associated files available for download.
More information
Published date: May 2009
Identifiers
Local EPrints ID: 163489
URI: http://eprints.soton.ac.uk/id/eprint/163489
ISSN: 0148-0731
PURE UUID: b6dceee6-00e5-4ac9-b545-ed05342b7f39
Catalogue record
Date deposited: 09 Sep 2010 07:53
Last modified: 02 Aug 2023 01:37
Export record
Altmetrics
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics