The reliability of a new measure of inspiratory muscle function in patients with COPD
The reliability of a new measure of inspiratory muscle function in patients with COPD
Rationale: Maximal inspiratory pressure (MIP) is an accepted effort-dependent measure of global inspiratory muscle strength. Prolonged MIP (PMIP) area is a new measure of inspiratory muscle function calculated as the area under the curve produced by sampling maximal inspiratory pressure at 16 Hz during a sustained inspiratory manoeuvre. The reliability of this measure has previously been established in normal subjects, but no published data are available for COPD patients.
Methods: Forty-three COPD patients attended for data collection on two occasions, 48 hours apart. Anthropometric data were collected on the first visit, pulmonary function, inspiratory muscle and peripheral muscle function (hand grip) data were collected during both visits. Calculated reliability estimates included the intraclass correlation coefficient (ICC), coefficient of variation (CV) and Bland & Altman 95% limits of agreement.
Results: Within-day reliability was consistently higher than between-day reliability. Between-day results generated an ICC of 0.92, a CV of 10% and 95% limits of agreement of - 58 to + 71 pressure time units. Hand grip data generated an ICC of 0.92 and a CV of 8%.
Conclusion: In this study group PMIP area was found to be a reproducible measure of inspiratory muscle function, generating comparable levels of reliability to those found in other effort-dependent tests of skeletal muscle function.
care
A156
Bruton, A.
a8cc274a-64af-4884-b76b-618ef6c29dad
Conway, J.H.
31968645-34bf-49e2-906d-b3e3487f13e5
Holgate, S.T.
48e197d1-3681-43c1-9a3e-478311f6618d
2001
Bruton, A.
a8cc274a-64af-4884-b76b-618ef6c29dad
Conway, J.H.
31968645-34bf-49e2-906d-b3e3487f13e5
Holgate, S.T.
48e197d1-3681-43c1-9a3e-478311f6618d
Bruton, A., Conway, J.H. and Holgate, S.T.
(2001)
The reliability of a new measure of inspiratory muscle function in patients with COPD.
American Journal of Respiratory and Critical Care Medicine, 163 (Supplement), .
Abstract
Rationale: Maximal inspiratory pressure (MIP) is an accepted effort-dependent measure of global inspiratory muscle strength. Prolonged MIP (PMIP) area is a new measure of inspiratory muscle function calculated as the area under the curve produced by sampling maximal inspiratory pressure at 16 Hz during a sustained inspiratory manoeuvre. The reliability of this measure has previously been established in normal subjects, but no published data are available for COPD patients.
Methods: Forty-three COPD patients attended for data collection on two occasions, 48 hours apart. Anthropometric data were collected on the first visit, pulmonary function, inspiratory muscle and peripheral muscle function (hand grip) data were collected during both visits. Calculated reliability estimates included the intraclass correlation coefficient (ICC), coefficient of variation (CV) and Bland & Altman 95% limits of agreement.
Results: Within-day reliability was consistently higher than between-day reliability. Between-day results generated an ICC of 0.92, a CV of 10% and 95% limits of agreement of - 58 to + 71 pressure time units. Hand grip data generated an ICC of 0.92 and a CV of 8%.
Conclusion: In this study group PMIP area was found to be a reproducible measure of inspiratory muscle function, generating comparable levels of reliability to those found in other effort-dependent tests of skeletal muscle function.
This record has no associated files available for download.
More information
Published date: 2001
Additional Information:
Poster presentation A46, B81 at 2001 Meeting of the American Thoracic Society, San Francisco, USA, 18-23 May 2001
Keywords:
care
Identifiers
Local EPrints ID: 17839
URI: http://eprints.soton.ac.uk/id/eprint/17839
PURE UUID: ea67680f-724d-41d9-adcb-298c75cff136
Catalogue record
Date deposited: 30 May 2006
Last modified: 11 Dec 2021 14:16
Export record
Contributors
Author:
A. Bruton
Author:
J.H. Conway
Author:
S.T. Holgate