Geometry parameterization and multidisciplinary constrained optimization of coronary stents

Geometry parameterization and multidisciplinary constrained optimization of coronary stents
Geometry parameterization and multidisciplinary constrained optimization of coronary stents
Coronary stents are tubular type scaffolds that are deployed, using an inflatable balloon on a catheter, most commonly to recover the lumen size of narrowed (diseased) arterial segments. A common differentiating factor between the numerous stents used in clinical practice today is their geometric design. An ideal stent should have high radial strength to provide good arterial support post-expansion, have high flexibility for easy manoeuvrability during deployment, cause minimal injury to the artery when being expanded and, for drug eluting stents, should provide adequate drug in the arterial tissue. Often, with any stent design, these objectives are in competition such that improvement in one objective is a result of trade-off in others. This study proposes a technique to parameterize stent geometry, by varying the shape of circumferential rings and the links, and assess performance by modelling the processes of balloon expansion and drug diffusion. Finite element analysis is used to expand each stent (through balloon inflation) into contact with a representative diseased coronary artery model, followed by a drug release simulation. Also, a separate model is constructed to measure stent flexibility. Since the computational simulation time for each design is very high (approximately 24 h), a Gaussian process modelling approach is used to analyse the design space corresponding to the proposed parameterization. Four objectives to assess recoil, stress distribution, drug distribution and flexibility are set up to perform optimization studies. In particular, single objective constrained optimization problems are set up to improve the design relative to the baseline geometry—i.e. to improve one objective without compromising the others. Improvements of 8, 6 and 15% are obtained individually for stress, drug and flexibility metrics, respectively. The relative influence of the design features on each objective is quantified in terms of main effects, thereby suggesting the design features which could be altered to improve stent performance. In particular, it is shown that large values of strut width combined with smaller axial lengths of circumferential rings are optimal in terms of minimizing average stresses and maximizing drug delivery. Furthermore, it is shown that a larger amplitude of the links with minimum curved regions is desirable for improved flexibility, average stresses and drug delivery
1617-7959
61-82
Pant, Sanjay
025d5228-8ee8-4269-96e8-0558c55a5f61
Bressloff, Neil W.
4f531e64-dbb3-41e3-a5d3-e6a5a7a77c92
Limbert, Georges
a1b88cb4-c5d9-4c6e-b6c9-7f4c4aa1c2ec
Pant, Sanjay
025d5228-8ee8-4269-96e8-0558c55a5f61
Bressloff, Neil W.
4f531e64-dbb3-41e3-a5d3-e6a5a7a77c92
Limbert, Georges
a1b88cb4-c5d9-4c6e-b6c9-7f4c4aa1c2ec

Pant, Sanjay, Bressloff, Neil W. and Limbert, Georges (2012) Geometry parameterization and multidisciplinary constrained optimization of coronary stents. Biomechanics and Modeling in Mechanobiology, 11 (1-2), 61-82. (doi:10.1007/s10237-011-0293-3).

Record type: Article

Abstract

Coronary stents are tubular type scaffolds that are deployed, using an inflatable balloon on a catheter, most commonly to recover the lumen size of narrowed (diseased) arterial segments. A common differentiating factor between the numerous stents used in clinical practice today is their geometric design. An ideal stent should have high radial strength to provide good arterial support post-expansion, have high flexibility for easy manoeuvrability during deployment, cause minimal injury to the artery when being expanded and, for drug eluting stents, should provide adequate drug in the arterial tissue. Often, with any stent design, these objectives are in competition such that improvement in one objective is a result of trade-off in others. This study proposes a technique to parameterize stent geometry, by varying the shape of circumferential rings and the links, and assess performance by modelling the processes of balloon expansion and drug diffusion. Finite element analysis is used to expand each stent (through balloon inflation) into contact with a representative diseased coronary artery model, followed by a drug release simulation. Also, a separate model is constructed to measure stent flexibility. Since the computational simulation time for each design is very high (approximately 24 h), a Gaussian process modelling approach is used to analyse the design space corresponding to the proposed parameterization. Four objectives to assess recoil, stress distribution, drug distribution and flexibility are set up to perform optimization studies. In particular, single objective constrained optimization problems are set up to improve the design relative to the baseline geometry—i.e. to improve one objective without compromising the others. Improvements of 8, 6 and 15% are obtained individually for stress, drug and flexibility metrics, respectively. The relative influence of the design features on each objective is quantified in terms of main effects, thereby suggesting the design features which could be altered to improve stent performance. In particular, it is shown that large values of strut width combined with smaller axial lengths of circumferential rings are optimal in terms of minimizing average stresses and maximizing drug delivery. Furthermore, it is shown that a larger amplitude of the links with minimum curved regions is desirable for improved flexibility, average stresses and drug delivery

Text
BMMB-Online-March2011].pdf - Version of Record
Restricted to Repository staff only
Request a copy

More information

Published date: 2012
Organisations: Computational Engineering and Design, Bioengineering Sciences, Engineering Science Unit

Identifiers

Local EPrints ID: 179759
URI: http://eprints.soton.ac.uk/id/eprint/179759
ISSN: 1617-7959
PURE UUID: 2f8f73b7-a8f1-4294-9eb1-4e753ac82b7e

Catalogue record

Date deposited: 06 May 2011 08:29
Last modified: 14 Mar 2024 02:50

Export record

Altmetrics

Contributors

Author: Sanjay Pant
Author: Georges Limbert

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×