Early femoral loosening of the titanium 3M Capitol cemented hip : A lesson re-learned
Early femoral loosening of the titanium 3M Capitol cemented hip : A lesson re-learned
A 4-year follow-up study is presented comparing 3M Capital cemented hip replacements with Charnley low friction arthroplasty controls. The mean radiological follow-up in the Capital and Charnley groups was 47 months (25 to 61) and 47 months (23 to 66) respectively. All cases were performed by the same surgical team over the same period (April 1991-October 1993). The operative technique including femoral preparation and cementing technique was identical in each group. Charnley broaches were used for femoral preparation in both groups and Charnley cups were used in all cases. In all of the 3m Capital cases cobalt-chrome femoral heads were used. There was no case selection for either group. Six of the 21 Capital femoral components were deemed to be loose on radiological criteria. None of the Charnley hips was loose. This represents an unacceptably high incidence of radiological femoral stem loosening in the Capital hip replacement. A previous uncontrolled study has attributed early failures to deficient femoral preparation and/or cementing technique. We suggest that early femoral stem loosening in the Capital hip occurs despite satisfactory operative technique in contrast with the results using the Charnley total hip replacement. (Hip International 2002; 1: 17-22)
17-22
Harvey, A.R.
2798b425-bc3b-46ac-9896-1d8a71baf509
Barlow, L.W.
371aafb6-b079-40c5-85e9-c94ad6b69ba3
Clarke, N.M.P.
76688c21-d51e-48fa-a84d-deec66baf8ac
2002
Harvey, A.R.
2798b425-bc3b-46ac-9896-1d8a71baf509
Barlow, L.W.
371aafb6-b079-40c5-85e9-c94ad6b69ba3
Clarke, N.M.P.
76688c21-d51e-48fa-a84d-deec66baf8ac
Harvey, A.R., Barlow, L.W. and Clarke, N.M.P.
(2002)
Early femoral loosening of the titanium 3M Capitol cemented hip : A lesson re-learned.
Hip International, 12 (1), .
Abstract
A 4-year follow-up study is presented comparing 3M Capital cemented hip replacements with Charnley low friction arthroplasty controls. The mean radiological follow-up in the Capital and Charnley groups was 47 months (25 to 61) and 47 months (23 to 66) respectively. All cases were performed by the same surgical team over the same period (April 1991-October 1993). The operative technique including femoral preparation and cementing technique was identical in each group. Charnley broaches were used for femoral preparation in both groups and Charnley cups were used in all cases. In all of the 3m Capital cases cobalt-chrome femoral heads were used. There was no case selection for either group. Six of the 21 Capital femoral components were deemed to be loose on radiological criteria. None of the Charnley hips was loose. This represents an unacceptably high incidence of radiological femoral stem loosening in the Capital hip replacement. A previous uncontrolled study has attributed early failures to deficient femoral preparation and/or cementing technique. We suggest that early femoral stem loosening in the Capital hip occurs despite satisfactory operative technique in contrast with the results using the Charnley total hip replacement. (Hip International 2002; 1: 17-22)
This record has no associated files available for download.
More information
Published date: 2002
Identifiers
Local EPrints ID: 190501
URI: http://eprints.soton.ac.uk/id/eprint/190501
ISSN: 1120-7000
PURE UUID: b109f7cf-ce99-40c6-a259-71ad998b5600
Catalogue record
Date deposited: 13 Jun 2011 14:23
Last modified: 08 Jan 2022 02:33
Export record
Contributors
Author:
A.R. Harvey
Author:
L.W. Barlow
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics