The University of Southampton
University of Southampton Institutional Repository

Temporal Optimization of Microfluidic Colorimetric Sensors by Use of Multiplexed Stop-Flow Architecture

Ogilvie, I.R.G., Sieben, V.J., Mowlem, M.C. and Morgan, H. (2011) Temporal Optimization of Microfluidic Colorimetric Sensors by Use of Multiplexed Stop-Flow Architecture Analytical Chemistry, 83, (12), pp. 4814-4821. (doi:10.1021/ac200463y).

Record type: Article

Abstract

We present two microfluidic architectures (continuous flow and multiplexed stop flow) for miniaturized colorimetric nutrient sensors. These systems are compared with respect to the temporal response (for optimization of sampling rate) and reduction of reagent consumption. The continuous-flow system is capable of a sampling rate of 60 samples·h–1, limited by Taylor dispersion. The novel multiplexed stop-flow (MSF) microsystem architecture is not limited by dispersion. A demonstration MSF system consisting of two stop-flow channels is presented. This requires 12.6 s to load each sample into a measurement channel and when scaled would be capable of a throughput of 285 h–1 (with full color development). The MSF architecture is manufactured in PMMA/Viton/PMMA [where PMMA = poly(methyl methacrylate)], utilizes on-chip valving, and is scalable, thereby permitting sampling at much faster rates (subsecond). Either system is capable of remote deployment and continuous measurement of nutrient concentrations. The MSF system is particularly suited for applications requiring high temporal or spatial resolution; such as from moving vehicles.

Full text not available from this repository.

More information

Published date: 2011
Organisations: Ocean Technology and Engineering

Identifiers

Local EPrints ID: 192321
URI: http://eprints.soton.ac.uk/id/eprint/192321
ISSN: 0003-2700
PURE UUID: 8b771cbc-a2e7-4f27-84ea-42fd0e3d39bc
ORCID for H. Morgan: ORCID iD orcid.org/0000-0003-4850-5676

Catalogue record

Date deposited: 01 Jul 2011 13:32
Last modified: 18 Jul 2017 11:32

Export record

Altmetrics

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×