Explicit finite element simulation of eccentric loading in total knee replacement
Explicit finite element simulation of eccentric loading in total knee replacement
In vivo kinematic data indicate that unicondylar or edge loading occurs during normal activities in well-aligned and malpositioned knee replacements. Using a validated explicit finite element model of a knee replacement, the effects of eccentric loading of a total knee replacement are simulated. Only minor variations were observed in the kinematics with a medial offset of the vertical load of as much as 15 mm (representing a medial:lateral loading ratio of 86:14), although the polyethylene stresses did increase by approximately 3 MPa throughout the stance phase of gait. There was a significant change in the kinematics and stresses when unicondylar loading occurred (95:5 medial:lateral loading ratio). Even for the unicondylar load case, contact always was maintained within the lateral compartment. This raises the question whether lift-off often observed in fluoroscopy studies really occurs. The model predicted regions of plastic deformation that closely resemble those observed in retrieved specimens of catastrophic wear. The explicit finite element model offers considerable insight into the kinematics and stresses generated by total knee replacement during different and varied loading conditions that occur during normal usage.
162-171
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
Barrett, D.S.
cbaaf354-cc75-42d4-980c-0a8797db5b4c
2003
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
Barrett, D.S.
cbaaf354-cc75-42d4-980c-0a8797db5b4c
Taylor, M. and Barrett, D.S.
(2003)
Explicit finite element simulation of eccentric loading in total knee replacement.
Clinical Orthopaedics and Related Research, 414 (162-171), .
(doi:10.1097/01.blo.0000079267.91782.58).
Abstract
In vivo kinematic data indicate that unicondylar or edge loading occurs during normal activities in well-aligned and malpositioned knee replacements. Using a validated explicit finite element model of a knee replacement, the effects of eccentric loading of a total knee replacement are simulated. Only minor variations were observed in the kinematics with a medial offset of the vertical load of as much as 15 mm (representing a medial:lateral loading ratio of 86:14), although the polyethylene stresses did increase by approximately 3 MPa throughout the stance phase of gait. There was a significant change in the kinematics and stresses when unicondylar loading occurred (95:5 medial:lateral loading ratio). Even for the unicondylar load case, contact always was maintained within the lateral compartment. This raises the question whether lift-off often observed in fluoroscopy studies really occurs. The model predicted regions of plastic deformation that closely resemble those observed in retrieved specimens of catastrophic wear. The explicit finite element model offers considerable insight into the kinematics and stresses generated by total knee replacement during different and varied loading conditions that occur during normal usage.
Text
00003086-200309000-00022.pdf
- Version of Record
Restricted to Repository staff only
Request a copy
More information
Published date: 2003
Identifiers
Local EPrints ID: 22547
URI: http://eprints.soton.ac.uk/id/eprint/22547
ISSN: 0009-921X
PURE UUID: 484d66e6-b80f-4e8b-af9c-0f777e34a5b8
Catalogue record
Date deposited: 27 Mar 2006
Last modified: 15 Mar 2024 06:38
Export record
Altmetrics
Contributors
Author:
M. Taylor
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics