A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration

A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration
A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration
There is strong evidence to suggest that inducible displacements, migration and implant loosening are closely related to the initial mechanical environment of the implanted tibia. If this is true, then it should be possible to predict the likelihood of implant migration using patient-specific finite element models. Finite element models of the proximal implanted tibiae were analysed based on pre-operative quantitative computed tomography data of four patients entered into a radiographic migration study. These four patients were also part of an radiostereometric analysis (RSA) study. A variety of load cases were analysed and the risk of bone failure determined for a 2 mm layer of bone immediately beneath the tibial tray. The results were compared with the RSA data measured 1 year post-operatively for each patient. For each patient, an appropriate load case was selected based on patient weight and on the varus–valgus migrations observed in the migration study. The two patients with press-fit implants were predicted to have the highest risk of failure and were found to migrate the most. The two patients with bonded implants (one HA coated and one cemented) were found to have a low risk of failure and these implants migrated the least.
This study suggests that the degree of implant migration is dependent on the initial mechanical environment and can be determined using patient-specific finite element analysis.
total knee replacement, finite element analysis, rsa
0021-9290
1205-1213
Perillo-Marcone, A.
14bf1408-6e29-46e1-9376-ecd9ae748814
Ryd, L.
4b08bb72-7753-4cfa-a79e-8756fd4b79f3
Johnsson, K.
67db95b0-cead-410e-bd5b-34fefbb13999
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
Perillo-Marcone, A.
14bf1408-6e29-46e1-9376-ecd9ae748814
Ryd, L.
4b08bb72-7753-4cfa-a79e-8756fd4b79f3
Johnsson, K.
67db95b0-cead-410e-bd5b-34fefbb13999
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb

Perillo-Marcone, A., Ryd, L., Johnsson, K. and Taylor, M. (2004) A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration. Journal of Biomechanics, 37 (8), 1205-1213. (doi:10.1016/j.jbiomech.2003.12.004).

Record type: Article

Abstract

There is strong evidence to suggest that inducible displacements, migration and implant loosening are closely related to the initial mechanical environment of the implanted tibia. If this is true, then it should be possible to predict the likelihood of implant migration using patient-specific finite element models. Finite element models of the proximal implanted tibiae were analysed based on pre-operative quantitative computed tomography data of four patients entered into a radiographic migration study. These four patients were also part of an radiostereometric analysis (RSA) study. A variety of load cases were analysed and the risk of bone failure determined for a 2 mm layer of bone immediately beneath the tibial tray. The results were compared with the RSA data measured 1 year post-operatively for each patient. For each patient, an appropriate load case was selected based on patient weight and on the varus–valgus migrations observed in the migration study. The two patients with press-fit implants were predicted to have the highest risk of failure and were found to migrate the most. The two patients with bonded implants (one HA coated and one cemented) were found to have a low risk of failure and these implants migrated the least.
This study suggests that the degree of implant migration is dependent on the initial mechanical environment and can be determined using patient-specific finite element analysis.

This record has no associated files available for download.

More information

Published date: 2004
Keywords: total knee replacement, finite element analysis, rsa
Organisations: Engineering Sciences

Identifiers

Local EPrints ID: 22847
URI: http://eprints.soton.ac.uk/id/eprint/22847
ISSN: 0021-9290
PURE UUID: 8ceb181f-65f8-45f8-9fa8-3af186f11fc5

Catalogue record

Date deposited: 23 Mar 2006
Last modified: 15 Mar 2024 06:41

Export record

Altmetrics

Contributors

Author: A. Perillo-Marcone
Author: L. Ryd
Author: K. Johnsson
Author: M. Taylor

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×