A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration
A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration
There is strong evidence to suggest that inducible displacements, migration and implant loosening are closely related to the initial mechanical environment of the implanted tibia. If this is true, then it should be possible to predict the likelihood of implant migration using patient-specific finite element models. Finite element models of the proximal implanted tibiae were analysed based on pre-operative quantitative computed tomography data of four patients entered into a radiographic migration study. These four patients were also part of an radiostereometric analysis (RSA) study. A variety of load cases were analysed and the risk of bone failure determined for a 2 mm layer of bone immediately beneath the tibial tray. The results were compared with the RSA data measured 1 year post-operatively for each patient. For each patient, an appropriate load case was selected based on patient weight and on the varus–valgus migrations observed in the migration study. The two patients with press-fit implants were predicted to have the highest risk of failure and were found to migrate the most. The two patients with bonded implants (one HA coated and one cemented) were found to have a low risk of failure and these implants migrated the least.
This study suggests that the degree of implant migration is dependent on the initial mechanical environment and can be determined using patient-specific finite element analysis.
total knee replacement, finite element analysis, rsa
1205-1213
Perillo-Marcone, A.
14bf1408-6e29-46e1-9376-ecd9ae748814
Ryd, L.
4b08bb72-7753-4cfa-a79e-8756fd4b79f3
Johnsson, K.
67db95b0-cead-410e-bd5b-34fefbb13999
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
2004
Perillo-Marcone, A.
14bf1408-6e29-46e1-9376-ecd9ae748814
Ryd, L.
4b08bb72-7753-4cfa-a79e-8756fd4b79f3
Johnsson, K.
67db95b0-cead-410e-bd5b-34fefbb13999
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
Perillo-Marcone, A., Ryd, L., Johnsson, K. and Taylor, M.
(2004)
A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration.
Journal of Biomechanics, 37 (8), .
(doi:10.1016/j.jbiomech.2003.12.004).
Abstract
There is strong evidence to suggest that inducible displacements, migration and implant loosening are closely related to the initial mechanical environment of the implanted tibia. If this is true, then it should be possible to predict the likelihood of implant migration using patient-specific finite element models. Finite element models of the proximal implanted tibiae were analysed based on pre-operative quantitative computed tomography data of four patients entered into a radiographic migration study. These four patients were also part of an radiostereometric analysis (RSA) study. A variety of load cases were analysed and the risk of bone failure determined for a 2 mm layer of bone immediately beneath the tibial tray. The results were compared with the RSA data measured 1 year post-operatively for each patient. For each patient, an appropriate load case was selected based on patient weight and on the varus–valgus migrations observed in the migration study. The two patients with press-fit implants were predicted to have the highest risk of failure and were found to migrate the most. The two patients with bonded implants (one HA coated and one cemented) were found to have a low risk of failure and these implants migrated the least.
This study suggests that the degree of implant migration is dependent on the initial mechanical environment and can be determined using patient-specific finite element analysis.
This record has no associated files available for download.
More information
Published date: 2004
Keywords:
total knee replacement, finite element analysis, rsa
Organisations:
Engineering Sciences
Identifiers
Local EPrints ID: 22847
URI: http://eprints.soton.ac.uk/id/eprint/22847
ISSN: 0021-9290
PURE UUID: 8ceb181f-65f8-45f8-9fa8-3af186f11fc5
Catalogue record
Date deposited: 23 Mar 2006
Last modified: 15 Mar 2024 06:41
Export record
Altmetrics
Contributors
Author:
A. Perillo-Marcone
Author:
L. Ryd
Author:
K. Johnsson
Author:
M. Taylor
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics