Quality control of slope-intercept measurements of glomerular filtration rate using single-sample estimates

Quality control of slope-intercept measurements of glomerular filtration rate using single-sample estimates
Quality control of slope-intercept measurements of glomerular filtration rate using single-sample estimates
Objectives: Measurement of glomerular filtration rate (GFR) using the slope-intercept technique determines the plasma clearance curve by fitting a straight line to the logarithm of sample count rate. When two samples are used there is no check on the validity of curve fitting. GFR may also be estimated from single-sample concentrations. This study describes a method of quality control for the two-sample technique using the agreement between the one-sample and two-sample estimates.
Methods: GFR measurements using 99mTc-DTPA were performed on 225 adults and 100 children using two samples taken between 2 h and 4 h post-injection. The two-sample values obtained using the British Nuclear Medicine Guidelines slope-intercept technique were compared to one-sample estimates obtained using a new general equation. Equations describing the variation of GFR error with GFR value were defined. These were used to determine action levels giving the limits of expected agreement between slope-intercept and single-sample values. The use of these action levels for quality control was demonstrated in a further 120 GFR measurements.
Results: The variation of single-sample error estimate with GFR depended both on the time of sample and body surface area. For specific sample groups, the error variation with GFR could be approximated using a truncated quadratic equation. Four studies were identified as failing quality control in the dataset used to define the error equations. Two studies failed in the test dataset.
Conclusions: One-sample equations give reliable estimates of GFR, which may be used for quality control of slope-intercept GFR assessment.
737-742
Fleming, John S.
9a23102d-6740-430a-ac74-89938b328ec3
Persaud, Linda
3d1f24dd-1d68-4a57-961e-6aef6ff22935
Zivanovic, Maureen A.
aff9ab8d-a09a-46c7-b70f-89a4a65695ed
Fleming, John S.
9a23102d-6740-430a-ac74-89938b328ec3
Persaud, Linda
3d1f24dd-1d68-4a57-961e-6aef6ff22935
Zivanovic, Maureen A.
aff9ab8d-a09a-46c7-b70f-89a4a65695ed

Fleming, John S., Persaud, Linda and Zivanovic, Maureen A. (2005) Quality control of slope-intercept measurements of glomerular filtration rate using single-sample estimates. Nuclear Medicine Communications, 26 (8), 737-742.

Record type: Article

Abstract

Objectives: Measurement of glomerular filtration rate (GFR) using the slope-intercept technique determines the plasma clearance curve by fitting a straight line to the logarithm of sample count rate. When two samples are used there is no check on the validity of curve fitting. GFR may also be estimated from single-sample concentrations. This study describes a method of quality control for the two-sample technique using the agreement between the one-sample and two-sample estimates.
Methods: GFR measurements using 99mTc-DTPA were performed on 225 adults and 100 children using two samples taken between 2 h and 4 h post-injection. The two-sample values obtained using the British Nuclear Medicine Guidelines slope-intercept technique were compared to one-sample estimates obtained using a new general equation. Equations describing the variation of GFR error with GFR value were defined. These were used to determine action levels giving the limits of expected agreement between slope-intercept and single-sample values. The use of these action levels for quality control was demonstrated in a further 120 GFR measurements.
Results: The variation of single-sample error estimate with GFR depended both on the time of sample and body surface area. For specific sample groups, the error variation with GFR could be approximated using a truncated quadratic equation. Four studies were identified as failing quality control in the dataset used to define the error equations. Two studies failed in the test dataset.
Conclusions: One-sample equations give reliable estimates of GFR, which may be used for quality control of slope-intercept GFR assessment.

This record has no associated files available for download.

More information

Published date: 2005

Identifiers

Local EPrints ID: 25499
URI: http://eprints.soton.ac.uk/id/eprint/25499
PURE UUID: a389d082-6f04-4de9-a5ef-16c9017388b8

Catalogue record

Date deposited: 10 Apr 2006
Last modified: 27 Apr 2022 05:20

Export record

Contributors

Author: John S. Fleming
Author: Linda Persaud
Author: Maureen A. Zivanovic

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×