The purpose of this study was to perform an audit of quantitative values obtained from gamma camera renography in the UK. Ten patient image sequences representing normal and pathological renal function were obtained from archived studies and distributed to hospitals in the UK. Hospitals were asked to measure five parameters: relative function, renogram time-to-peak (left and right), and whole kidney mean transit time (left and right). Details of methodology, software used and operator experience were requested. This allowed the influence of operational factors on variations in reported values to be examined. A total of 180 responses from 81 hospitals were received. Values reported for the parameters, together with other details supplied, were entered into Excel and SPSS for statistical analysis. Histograms representing the distribution of values were produced for each parameter. The largest variations were found for mean transit time and occasionally for time-to-peak. The effect of factors was assessed using non-parametric statistical tests applied independently to each renogram. For all the parameters, the hospital, UK region, supplier, computer and software version influenced variations in the reported values. Algorithm and site of background region were influencing factors for relative function, the background subtraction method influenced time-to-peak, and curve smoothing influenced mean transit time.
559-566
Houston, A.S.
8007585b-7ac5-4236-a1f2-8b4c34ca6d6e
Whalley, D.R.
e374f7ae-a3bc-42ee-8684-ef172e6cf313
Skrypniuk, J.V.
7680451a-99c9-4bae-b13b-c9251ee354d1
Jarritt, P.H.
e57bd842-8104-421f-8951-e30cd7617920
Fleming, J.S.
bf089a52-bc49-402a-9129-0dfecf452936
Cosgriff, P.S.
2d3cb87f-b58b-472c-b2d3-80c3a82ba44e
Houston, A.S.
8007585b-7ac5-4236-a1f2-8b4c34ca6d6e
Whalley, D.R.
e374f7ae-a3bc-42ee-8684-ef172e6cf313
Skrypniuk, J.V.
7680451a-99c9-4bae-b13b-c9251ee354d1
Jarritt, P.H.
e57bd842-8104-421f-8951-e30cd7617920
Fleming, J.S.
bf089a52-bc49-402a-9129-0dfecf452936
Cosgriff, P.S.
2d3cb87f-b58b-472c-b2d3-80c3a82ba44e

Houston, A.S., Whalley, D.R., Skrypniuk, J.V., Jarritt, P.H., Fleming, J.S. and Cosgriff, P.S. (2001) UK audit and analysis of quantitative parameters obtained from gamma camera renography. Nuclear Medicine Communications, 22 (5), 559-566.

Record type: Article

Abstract

The purpose of this study was to perform an audit of quantitative values obtained from gamma camera renography in the UK. Ten patient image sequences representing normal and pathological renal function were obtained from archived studies and distributed to hospitals in the UK. Hospitals were asked to measure five parameters: relative function, renogram time-to-peak (left and right), and whole kidney mean transit time (left and right). Details of methodology, software used and operator experience were requested. This allowed the influence of operational factors on variations in reported values to be examined. A total of 180 responses from 81 hospitals were received. Values reported for the parameters, together with other details supplied, were entered into Excel and SPSS for statistical analysis. Histograms representing the distribution of values were produced for each parameter. The largest variations were found for mean transit time and occasionally for time-to-peak. The effect of factors was assessed using non-parametric statistical tests applied independently to each renogram. For all the parameters, the hospital, UK region, supplier, computer and software version influenced variations in the reported values. Algorithm and site of background region were influencing factors for relative function, the background subtraction method influenced time-to-peak, and curve smoothing influenced mean transit time.

This record has no associated files available for download.

More information

Published date: 2001

Identifiers

Local EPrints ID: 25642
URI: http://eprints.soton.ac.uk/id/eprint/25642
PURE UUID: e7b6c716-ba46-4ce0-aa1d-90fc82473568

Catalogue record

Date deposited: 11 Apr 2006
Last modified: 08 Jan 2022 12:52

Export record

Contributors

Author: A.S. Houston
Author: D.R. Whalley
Author: J.V. Skrypniuk
Author: P.H. Jarritt
Author: J.S. Fleming
Author: P.S. Cosgriff

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×