Three-dimensional ultrasound fetal lung volume measurement: a systematic study comparing the multiplanar method with the rotational (VOCAL) technique
Three-dimensional ultrasound fetal lung volume measurement: a systematic study comparing the multiplanar method with the rotational (VOCAL) technique
Objectives:
This study was designed to compare a conventional multiplanar technique for three-dimensional (3D) ultrasound measurement of fetal lung volume with a rotational method using VOCALTM (Virtual Organ Computer-aided AnaLysis).
Methods:
Thirty-two fetuses with a variety of conditions at risk for pulmonary hypoplasia were studied. 3D volume data sets of the fetal lungs were acquired using a commercially available ultrasound system. The right and left lung volumes were calculated separately using VOCAL and the multiplanar technique. The level of agreement between two independent observers in categorizing the 3D volume data set as measurable or non-measurable was determined. The interobserver and intermethod variabilities were also evaluated for both methods.
Results:
The intermethod variability was excellent (correlation r = 0.93 and r = 0.96 for the left and right lung, respectively), and there was substantial agreement between the results of both approaches (limits of agreement - 4.4 to 8.9 and - 3.4 to 4.8 mL for the right and left lung, respectively). Fetal lung estimation with VOCAL had a significantly higher interobserver variability than the multiplanar technique. Interobserver agreement in categorizing lung volume data sets as measurable or non-measurable was lower when VOCAL was used.
Conclusion:
Fetal lung volume measurements can be undertaken interchangeably using the multiplanar technique or the rotational method with VOCAL. However, the latter was less reproducible (lower degree of agreement and significantly higher interobserver variability) than the former.
lung volume, pulmonary hypoplasia, three-dimensional ultrasonography, VOCAL
111-118
Kalache, K.D.
528a50b8-6bcf-42f6-9fcf-87825ed44304
Espinoza, J.
f3779ca9-13cc-4c5e-bf64-6c1579db10df
Chaiworapongsa, T.
6e456b97-2742-466f-9a2d-a096d4ec3500
Londono, J.
70291314-a51b-478a-9b00-87c96ac66dfb
Schoen, M.L.
91b7a9b3-250e-41f0-8086-c803c012b47b
Treadwell, M.C.
4be4aaa4-9e3a-49ab-a9ae-2fcc38cdefb1
Lee, W.
9e508846-252a-4bfe-bfff-d10be6c59ed9
Romero, R.
aa0ac4a7-b0d8-412d-9407-ffdaa76bda31
2003
Kalache, K.D.
528a50b8-6bcf-42f6-9fcf-87825ed44304
Espinoza, J.
f3779ca9-13cc-4c5e-bf64-6c1579db10df
Chaiworapongsa, T.
6e456b97-2742-466f-9a2d-a096d4ec3500
Londono, J.
70291314-a51b-478a-9b00-87c96ac66dfb
Schoen, M.L.
91b7a9b3-250e-41f0-8086-c803c012b47b
Treadwell, M.C.
4be4aaa4-9e3a-49ab-a9ae-2fcc38cdefb1
Lee, W.
9e508846-252a-4bfe-bfff-d10be6c59ed9
Romero, R.
aa0ac4a7-b0d8-412d-9407-ffdaa76bda31
Kalache, K.D., Espinoza, J., Chaiworapongsa, T., Londono, J., Schoen, M.L., Treadwell, M.C., Lee, W. and Romero, R.
(2003)
Three-dimensional ultrasound fetal lung volume measurement: a systematic study comparing the multiplanar method with the rotational (VOCAL) technique.
Ultrasound in Obstetrics & Gynecology, 21 (2), .
(doi:10.1002/uog.39).
Abstract
Objectives:
This study was designed to compare a conventional multiplanar technique for three-dimensional (3D) ultrasound measurement of fetal lung volume with a rotational method using VOCALTM (Virtual Organ Computer-aided AnaLysis).
Methods:
Thirty-two fetuses with a variety of conditions at risk for pulmonary hypoplasia were studied. 3D volume data sets of the fetal lungs were acquired using a commercially available ultrasound system. The right and left lung volumes were calculated separately using VOCAL and the multiplanar technique. The level of agreement between two independent observers in categorizing the 3D volume data set as measurable or non-measurable was determined. The interobserver and intermethod variabilities were also evaluated for both methods.
Results:
The intermethod variability was excellent (correlation r = 0.93 and r = 0.96 for the left and right lung, respectively), and there was substantial agreement between the results of both approaches (limits of agreement - 4.4 to 8.9 and - 3.4 to 4.8 mL for the right and left lung, respectively). Fetal lung estimation with VOCAL had a significantly higher interobserver variability than the multiplanar technique. Interobserver agreement in categorizing lung volume data sets as measurable or non-measurable was lower when VOCAL was used.
Conclusion:
Fetal lung volume measurements can be undertaken interchangeably using the multiplanar technique or the rotational method with VOCAL. However, the latter was less reproducible (lower degree of agreement and significantly higher interobserver variability) than the former.
This record has no associated files available for download.
More information
Published date: 2003
Keywords:
lung volume, pulmonary hypoplasia, three-dimensional ultrasonography, VOCAL
Identifiers
Local EPrints ID: 25700
URI: http://eprints.soton.ac.uk/id/eprint/25700
ISSN: 0960-7692
PURE UUID: 91b2d753-a0f2-4041-89eb-d5c6e0a7f036
Catalogue record
Date deposited: 19 Apr 2006
Last modified: 15 Mar 2024 07:04
Export record
Altmetrics
Contributors
Author:
K.D. Kalache
Author:
J. Espinoza
Author:
T. Chaiworapongsa
Author:
J. Londono
Author:
M.L. Schoen
Author:
M.C. Treadwell
Author:
W. Lee
Author:
R. Romero
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics