UK audit of left ventricular ejection fraction estimation from equilibrium ECG gated blood pool images
UK audit of left ventricular ejection fraction estimation from equilibrium ECG gated blood pool images
Purpose: To examine the variability of results obtained from computer analysis of left ventricular gated blood pool (LVGBP) images by nuclear medicine centres in the UK.
Methods: Twelve data sets of LVGBP images were distributed via commercial software suppliers to nuclear medicine centres in the UK. Two of the data sets were duplicates and three were acquired from the same patient with different total counts in the images. The quality of the images was also variable and two images had poorly defined left ventricular walls. A questionnaire was used to identify the parameters used during the analysis and to give an indication of the number of LVGBP scans per year routinely carried out by each centre as well as report the results obtained from the analysis.
Results: Results were received from 63 nuclear medicine centres using 77 computer systems. The vast majority of participants (57) carried out fewer than 10 scans per month. Only two centres performed more than 30 scans per month. Sixteen centres did not quote a minimum normal value for left ventricular ejection fraction (LVEF) and 36 did not record a maximum value. The remainder recorded between 0.40 and 0.60 for the minimum of normal range and 0.60-0.90 for the maximum of normal range. Analysis of returns showed that LVEF estimates for the data sets were highly variable between centres and computer systems. The overall standard deviation of results compared to the mean for each study was 0.076. Approximately half this variation was due to systematic variation between centres. The overall precision taking into consideration this systematic variation, was 0.055. Lower variability was found between studies with higher overall counts and this was highly significant.
205-215
Skrypniuk, J.V.
7680451a-99c9-4bae-b13b-c9251ee354d1
Bailey, D.
2b9988b1-d7f4-43a5-87a6-f61f7b287f32
Cosgriff, P.S.
2d3cb87f-b58b-472c-b2d3-80c3a82ba44e
Fleming, J.S.
bf089a52-bc49-402a-9129-0dfecf452936
Houston, A.S.
8007585b-7ac5-4236-a1f2-8b4c34ca6d6e
Jarritt, P.H.
e57bd842-8104-421f-8951-e30cd7617920
Whalley, D.R.
e374f7ae-a3bc-42ee-8684-ef172e6cf313
2005
Skrypniuk, J.V.
7680451a-99c9-4bae-b13b-c9251ee354d1
Bailey, D.
2b9988b1-d7f4-43a5-87a6-f61f7b287f32
Cosgriff, P.S.
2d3cb87f-b58b-472c-b2d3-80c3a82ba44e
Fleming, J.S.
bf089a52-bc49-402a-9129-0dfecf452936
Houston, A.S.
8007585b-7ac5-4236-a1f2-8b4c34ca6d6e
Jarritt, P.H.
e57bd842-8104-421f-8951-e30cd7617920
Whalley, D.R.
e374f7ae-a3bc-42ee-8684-ef172e6cf313
Skrypniuk, J.V., Bailey, D., Cosgriff, P.S., Fleming, J.S., Houston, A.S., Jarritt, P.H. and Whalley, D.R.
(2005)
UK audit of left ventricular ejection fraction estimation from equilibrium ECG gated blood pool images.
Nuclear Medicine Communications, 26 (3), .
Abstract
Purpose: To examine the variability of results obtained from computer analysis of left ventricular gated blood pool (LVGBP) images by nuclear medicine centres in the UK.
Methods: Twelve data sets of LVGBP images were distributed via commercial software suppliers to nuclear medicine centres in the UK. Two of the data sets were duplicates and three were acquired from the same patient with different total counts in the images. The quality of the images was also variable and two images had poorly defined left ventricular walls. A questionnaire was used to identify the parameters used during the analysis and to give an indication of the number of LVGBP scans per year routinely carried out by each centre as well as report the results obtained from the analysis.
Results: Results were received from 63 nuclear medicine centres using 77 computer systems. The vast majority of participants (57) carried out fewer than 10 scans per month. Only two centres performed more than 30 scans per month. Sixteen centres did not quote a minimum normal value for left ventricular ejection fraction (LVEF) and 36 did not record a maximum value. The remainder recorded between 0.40 and 0.60 for the minimum of normal range and 0.60-0.90 for the maximum of normal range. Analysis of returns showed that LVEF estimates for the data sets were highly variable between centres and computer systems. The overall standard deviation of results compared to the mean for each study was 0.076. Approximately half this variation was due to systematic variation between centres. The overall precision taking into consideration this systematic variation, was 0.055. Lower variability was found between studies with higher overall counts and this was highly significant.
This record has no associated files available for download.
More information
Published date: 2005
Identifiers
Local EPrints ID: 25983
URI: http://eprints.soton.ac.uk/id/eprint/25983
ISSN: 0143-3636
PURE UUID: 1e4e65af-a99b-42eb-a6ec-1dff35d1482e
Catalogue record
Date deposited: 20 Apr 2006
Last modified: 08 Jan 2022 18:55
Export record
Contributors
Author:
J.V. Skrypniuk
Author:
D. Bailey
Author:
P.S. Cosgriff
Author:
J.S. Fleming
Author:
A.S. Houston
Author:
P.H. Jarritt
Author:
D.R. Whalley
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics