Variation of young normal-hearing thresholds measured using different audiometric earphones: implications for the acoustic coupler and the ear simulator

Variation of young normal-hearing thresholds measured using different audiometric earphones: implications for the acoustic coupler and the ear simulator
Variation of young normal-hearing thresholds measured using different audiometric earphones: implications for the acoustic coupler and the ear simulator
This paper questions the necessity for two calibration devices to measure the acoustic output from different types of audiometric earphones. International standards give the audiometric zero for TDH39 earphones on the IEC 60318-3 acoustic coupler; the IEC 60318-1 ear simulator is intended for other supra-aural earphone types. If hearing threshold samples from young, healthy ears were found to be more variable using TDH39 earphones, then that earphone and its coupler might be taken out of service.
The audiological literature yielded threshold survey results for over 5100 otologically normal ears of subjects aged 31 years or less. These independent samples showed smaller variation for TDH39 samples than for samples using other earphones; this finding does not support abandoning the TDH39 and its coupler. Nevertheless, benefits accrue from calibrating TDH39 output to the audiometric zero as measured on the ear simulator.
hearing threshold, earphone, acoustic coupler, artificial ear, ear simulator
444-451
Lawton, Ben W.
4eb10a80-d554-4118-a0b2-68023b349f5e
Lawton, Ben W.
4eb10a80-d554-4118-a0b2-68023b349f5e

Lawton, Ben W. (2005) Variation of young normal-hearing thresholds measured using different audiometric earphones: implications for the acoustic coupler and the ear simulator. International Journal of Audiology, 44 (8), 444-451. (doi:10.1080/14992020500189062).

Record type: Article

Abstract

This paper questions the necessity for two calibration devices to measure the acoustic output from different types of audiometric earphones. International standards give the audiometric zero for TDH39 earphones on the IEC 60318-3 acoustic coupler; the IEC 60318-1 ear simulator is intended for other supra-aural earphone types. If hearing threshold samples from young, healthy ears were found to be more variable using TDH39 earphones, then that earphone and its coupler might be taken out of service.
The audiological literature yielded threshold survey results for over 5100 otologically normal ears of subjects aged 31 years or less. These independent samples showed smaller variation for TDH39 samples than for samples using other earphones; this finding does not support abandoning the TDH39 and its coupler. Nevertheless, benefits accrue from calibrating TDH39 output to the audiometric zero as measured on the ear simulator.

This record has no associated files available for download.

More information

Published date: 2005
Keywords: hearing threshold, earphone, acoustic coupler, artificial ear, ear simulator

Identifiers

Local EPrints ID: 28282
URI: http://eprints.soton.ac.uk/id/eprint/28282
PURE UUID: 2ed1294a-965b-459e-8abf-2471a9eaa460

Catalogue record

Date deposited: 02 May 2006
Last modified: 15 Mar 2024 07:23

Export record

Altmetrics

Contributors

Author: Ben W. Lawton

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×