Calibration of dual-beam spectroradiometric data
Calibration of dual-beam spectroradiometric data
An experiment to determine the most accurate and repeatable method for generating instrument inter-calibration functions (ICFs) is described, based upon data collected with a dual-beam GER1500 spectroradiometer system. The quality of reflectance data collected using a dual-beam spectroradiometer system is reliant upon accurate inter-calibration of the sensor pairs to take into account differences in their radiant sensitivity and spectral characteristics. A cos-conical field-based method for inter-calibrating dual-beam spectroradiometers was tested alongside laboratory inter-calibration procedures.
The field-based method produced the most accurate results when a field-derived ICF collected close in time was used to correct the spectral scan. A regression model to predict the ICF at a range of wavelengths was tested, using inputs of solar zenith angle, cosine of solar zenith angle and broadband diffuse-to-global irradiance ratios. The linear multiple regression model described up to 78% of the variability in ICF; the remainder of the variability was most likely due to complexities of instrumental
behaviour in response to warm-up time, ambient temperature and environmental conditions at the time of measurement. Collection of ICFs using a stable laboratory source was shown to provide unsatisfactory results due to differences between lamp outputs and the field-measured solar spectrum. Consequently, the most practical and accurate method of deriving inter-calibration functions is to use field-derived ICFs, collected close in time and space to the data requiring correction.
calibration, spectroradiometer, methodology
975-986
Anderson, K
6c2d56c2-26de-4a91-b79a-90324a1e4c9f
Milton, E.J.
f6cb5c0d-a5d4-47d7-860f-096de08e0c24
Rollin, E.M.
32665285-7f95-40a4-9775-7a9b2958f758
10 March 2006
Anderson, K
6c2d56c2-26de-4a91-b79a-90324a1e4c9f
Milton, E.J.
f6cb5c0d-a5d4-47d7-860f-096de08e0c24
Rollin, E.M.
32665285-7f95-40a4-9775-7a9b2958f758
Anderson, K, Milton, E.J. and Rollin, E.M.
(2006)
Calibration of dual-beam spectroradiometric data.
International Journal of Remote Sensing, 27 (5), .
(doi:10.1080/01431160500213375).
Abstract
An experiment to determine the most accurate and repeatable method for generating instrument inter-calibration functions (ICFs) is described, based upon data collected with a dual-beam GER1500 spectroradiometer system. The quality of reflectance data collected using a dual-beam spectroradiometer system is reliant upon accurate inter-calibration of the sensor pairs to take into account differences in their radiant sensitivity and spectral characteristics. A cos-conical field-based method for inter-calibrating dual-beam spectroradiometers was tested alongside laboratory inter-calibration procedures.
The field-based method produced the most accurate results when a field-derived ICF collected close in time was used to correct the spectral scan. A regression model to predict the ICF at a range of wavelengths was tested, using inputs of solar zenith angle, cosine of solar zenith angle and broadband diffuse-to-global irradiance ratios. The linear multiple regression model described up to 78% of the variability in ICF; the remainder of the variability was most likely due to complexities of instrumental
behaviour in response to warm-up time, ambient temperature and environmental conditions at the time of measurement. Collection of ICFs using a stable laboratory source was shown to provide unsatisfactory results due to differences between lamp outputs and the field-measured solar spectrum. Consequently, the most practical and accurate method of deriving inter-calibration functions is to use field-derived ICFs, collected close in time and space to the data requiring correction.
This record has no associated files available for download.
More information
Submitted date: 15 July 2004
Published date: 10 March 2006
Keywords:
calibration, spectroradiometer, methodology
Identifiers
Local EPrints ID: 28622
URI: http://eprints.soton.ac.uk/id/eprint/28622
ISSN: 0143-1161
PURE UUID: 18fb4ec1-9797-42b3-9884-e5ab3a3164fe
Catalogue record
Date deposited: 03 May 2006
Last modified: 15 Mar 2024 07:25
Export record
Altmetrics
Contributors
Author:
K Anderson
Author:
E.M. Rollin
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics