Calibration of dual-beam spectroradiometric data

Calibration of dual-beam spectroradiometric data
Calibration of dual-beam spectroradiometric data
An experiment to determine the most accurate and repeatable method for generating instrument inter-calibration functions (ICFs) is described, based upon data collected with a dual-beam GER1500 spectroradiometer system. The quality of reflectance data collected using a dual-beam spectroradiometer system is reliant upon accurate inter-calibration of the sensor pairs to take into account differences in their radiant sensitivity and spectral characteristics. A cos-conical field-based method for inter-calibrating dual-beam spectroradiometers was tested alongside laboratory inter-calibration procedures.
The field-based method produced the most accurate results when a field-derived ICF collected close in time was used to correct the spectral scan. A regression model to predict the ICF at a range of wavelengths was tested, using inputs of solar zenith angle, cosine of solar zenith angle and broadband diffuse-to-global irradiance ratios. The linear multiple regression model described up to 78% of the variability in ICF; the remainder of the variability was most likely due to complexities of instrumental behaviour in response to warm-up time, ambient temperature and environmental conditions at the time of measurement. Collection of ICFs using a stable laboratory source was shown to provide unsatisfactory results due to differences between lamp outputs and the field-measured solar spectrum. Consequently, the most practical and accurate method of deriving inter-calibration functions is to use field-derived ICFs, collected close in time and space to the data requiring correction.
calibration, spectroradiometer, methodology
0143-1161
975-986
Anderson, K
6c2d56c2-26de-4a91-b79a-90324a1e4c9f
Milton, E.J.
f6cb5c0d-a5d4-47d7-860f-096de08e0c24
Rollin, E.M.
32665285-7f95-40a4-9775-7a9b2958f758
Anderson, K
6c2d56c2-26de-4a91-b79a-90324a1e4c9f
Milton, E.J.
f6cb5c0d-a5d4-47d7-860f-096de08e0c24
Rollin, E.M.
32665285-7f95-40a4-9775-7a9b2958f758

Anderson, K, Milton, E.J. and Rollin, E.M. (2006) Calibration of dual-beam spectroradiometric data. International Journal of Remote Sensing, 27 (5), 975-986. (doi:10.1080/01431160500213375).

Record type: Article

Abstract

An experiment to determine the most accurate and repeatable method for generating instrument inter-calibration functions (ICFs) is described, based upon data collected with a dual-beam GER1500 spectroradiometer system. The quality of reflectance data collected using a dual-beam spectroradiometer system is reliant upon accurate inter-calibration of the sensor pairs to take into account differences in their radiant sensitivity and spectral characteristics. A cos-conical field-based method for inter-calibrating dual-beam spectroradiometers was tested alongside laboratory inter-calibration procedures.
The field-based method produced the most accurate results when a field-derived ICF collected close in time was used to correct the spectral scan. A regression model to predict the ICF at a range of wavelengths was tested, using inputs of solar zenith angle, cosine of solar zenith angle and broadband diffuse-to-global irradiance ratios. The linear multiple regression model described up to 78% of the variability in ICF; the remainder of the variability was most likely due to complexities of instrumental behaviour in response to warm-up time, ambient temperature and environmental conditions at the time of measurement. Collection of ICFs using a stable laboratory source was shown to provide unsatisfactory results due to differences between lamp outputs and the field-measured solar spectrum. Consequently, the most practical and accurate method of deriving inter-calibration functions is to use field-derived ICFs, collected close in time and space to the data requiring correction.

This record has no associated files available for download.

More information

Submitted date: 15 July 2004
Published date: 10 March 2006
Keywords: calibration, spectroradiometer, methodology

Identifiers

Local EPrints ID: 28622
URI: http://eprints.soton.ac.uk/id/eprint/28622
ISSN: 0143-1161
PURE UUID: 18fb4ec1-9797-42b3-9884-e5ab3a3164fe

Catalogue record

Date deposited: 03 May 2006
Last modified: 15 Mar 2024 07:25

Export record

Altmetrics

Contributors

Author: K Anderson
Author: E.J. Milton
Author: E.M. Rollin

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×