Investigation of the fabrication parameters of thick film oxide-PVC pH electrodes using experimental designs
Investigation of the fabrication parameters of thick film oxide-PVC pH electrodes using experimental designs
This paper describes a study into the development of an optimum fabrication process for the batch production of thick film titanium oxide-PVC pH electrodes. Multi-factorial experimental plans are used to explore the effect of fabrication parameters on the average sensitivity of sensors and their batch variability. Analysis of the results indicates that a number of the factors have a small effect on the sensitivity.
Factors of particular importance are identified and TiO2 in a PVC binder is examined in detail. The optimal values for these fabrication values are identified and examined. Mechanisms for the experimentally observed sub-Nernstian responses and immersion-time variations in sensor behaviour are discussed.
print media, optimization techniques, polymers, assembly, sensors, electrodes
44-53
Gac, A.
609d6dea-3257-4b3a-a4f5-80cf877a96a2
Atkinson, J.K.
5e9729b2-0e1f-400d-a889-c74f6390ea58
Zhang, Z.
fd968521-9e43-4b50-b67e-3b6dfa095d13
Sexton, C.J.
d8e295f8-6491-4952-942c-835d0343bce7
Lewis, S.M.
a69a3245-8c19-41c6-bf46-0b3b02d83cb8
Please, C.P.
118dffe7-4b38-4787-a972-9feec535839e
Sion, Russell
0fe05027-aae5-482f-b571-2bb9fc3491bf
2004
Gac, A.
609d6dea-3257-4b3a-a4f5-80cf877a96a2
Atkinson, J.K.
5e9729b2-0e1f-400d-a889-c74f6390ea58
Zhang, Z.
fd968521-9e43-4b50-b67e-3b6dfa095d13
Sexton, C.J.
d8e295f8-6491-4952-942c-835d0343bce7
Lewis, S.M.
a69a3245-8c19-41c6-bf46-0b3b02d83cb8
Please, C.P.
118dffe7-4b38-4787-a972-9feec535839e
Sion, Russell
0fe05027-aae5-482f-b571-2bb9fc3491bf
Gac, A., Atkinson, J.K., Zhang, Z., Sexton, C.J., Lewis, S.M., Please, C.P. and Sion, Russell
(2004)
Investigation of the fabrication parameters of thick film oxide-PVC pH electrodes using experimental designs.
Microelectronics International, 21 (3), .
(doi:10.1108/13565360410549710).
Abstract
This paper describes a study into the development of an optimum fabrication process for the batch production of thick film titanium oxide-PVC pH electrodes. Multi-factorial experimental plans are used to explore the effect of fabrication parameters on the average sensitivity of sensors and their batch variability. Analysis of the results indicates that a number of the factors have a small effect on the sensitivity.
Factors of particular importance are identified and TiO2 in a PVC binder is examined in detail. The optimal values for these fabrication values are identified and examined. Mechanisms for the experimentally observed sub-Nernstian responses and immersion-time variations in sensor behaviour are discussed.
This record has no associated files available for download.
More information
Published date: 2004
Keywords:
print media, optimization techniques, polymers, assembly, sensors, electrodes
Organisations:
Statistics
Identifiers
Local EPrints ID: 29265
URI: http://eprints.soton.ac.uk/id/eprint/29265
ISSN: 1356-5362
PURE UUID: 4e6faca4-e6e5-44d3-a562-08f0f07bdcdf
Catalogue record
Date deposited: 12 May 2006
Last modified: 16 Mar 2024 02:32
Export record
Altmetrics
Contributors
Author:
A. Gac
Author:
Z. Zhang
Author:
C.J. Sexton
Author:
S.M. Lewis
Author:
C.P. Please
Author:
Russell Sion
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics