The aim of this study was to use finite element models to investigate the effect of the design of the taper of polished, collarless, total hip replacement femoral components on stresses in the cement mantle surrounding the component. A single-taper prosthesis, double-taper prosthesis, and triple-taper prosthesis were compared. Peak stresses and stress distributions in the cement mantle were found to be a function of taper design, although the differences between designs were minor. Using a probability of failure technique based on the initial cement stress distribution, a triple-taper prosthesis was predicted to cause less cement mantle damage (0.15% of the volume of the cement mantle failing after 20 million loading cycles) than a double-taper prosthesis (0.74%) or a single-taper prosthesis (1.50%). Further research is required to confirm this finding.
s857-s862
New, A.M.R.
d2fbaf80-3abd-4bc5-ae36-9c77dfdde0d6
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
Wroblewski, B.M.
2e630ca9-55e6-44d2-9d6d-5bb7746c447b
August 2005
New, A.M.R.
d2fbaf80-3abd-4bc5-ae36-9c77dfdde0d6
Taylor, M.
e368bda3-6ca5-4178-80e9-41a689badeeb
Wroblewski, B.M.
2e630ca9-55e6-44d2-9d6d-5bb7746c447b
New, A.M.R., Taylor, M. and Wroblewski, B.M.
(2005)
Effect of hip stem taper on cement stresses.
Orthopedics, 28 (Supplement 8), .
Abstract
The aim of this study was to use finite element models to investigate the effect of the design of the taper of polished, collarless, total hip replacement femoral components on stresses in the cement mantle surrounding the component. A single-taper prosthesis, double-taper prosthesis, and triple-taper prosthesis were compared. Peak stresses and stress distributions in the cement mantle were found to be a function of taper design, although the differences between designs were minor. Using a probability of failure technique based on the initial cement stress distribution, a triple-taper prosthesis was predicted to cause less cement mantle damage (0.15% of the volume of the cement mantle failing after 20 million loading cycles) than a double-taper prosthesis (0.74%) or a single-taper prosthesis (1.50%). Further research is required to confirm this finding.
This record has no associated files available for download.
More information
Published date: August 2005
Identifiers
Local EPrints ID: 30132
URI: http://eprints.soton.ac.uk/id/eprint/30132
ISSN: 0147-7447
PURE UUID: bed8f3fd-cd4b-4736-ac11-322e060a9d29
Catalogue record
Date deposited: 11 May 2006
Last modified: 27 Apr 2022 08:45
Export record
Contributors
Author:
A.M.R. New
Author:
M. Taylor
Author:
B.M. Wroblewski
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics