Just scratching the surface: varied coagulation effects of polymer containers on TEG variables
Just scratching the surface: varied coagulation effects of polymer containers on TEG variables
Background and objective: Different types of polymer surfaces affect the activation of platelets and coagulation pathway containers depending on their surface qualities. Importantly, this could produce variability of coagulation results obtained with thrombelastographical analysis. We assessed the effects of blood storage on thrombelastograph, TEG, variables using polypropylene and polycarbonate containers.
Methods: An in vitro experiment was performed, with eight volunteers in each limb. Fresh whole blood was stored in polypropylene or polycarbonate tubes prior to TEGanalysis, to assess the role of these plastics in the TEG results obtained.
Results: The polycarbonate tubes displayed slower onset of coagulation and greater variability of data for all four basic TEG variables (r-time, k-time, alpha-angle and maximum amplitude, P < 0.05). Polycarbonate results fell outside manufacturer reference ranges.
Conclusions: It is likely that this is due to the altered surface properties and charge effects of the containers affecting proteins and platelets differently. Caution should be used in choosing which containers are used for storage of fresh blood prior to coagulation assessment, as variable results will follow where different types of plastic containers are employed.
45-49
Roche, A.M.
c14873d3-46ac-4df0-b178-b21e4835c8c4
James, M.F.M.
55614d48-1b93-4bbc-ac45-ccb28e0c384f
Grocott, M.P.W.
1e87b741-513e-4a22-be13-0f7bb344e8c2
Mythen, M.G.
c82ab585-86d5-44f8-bd3c-ead87c59d459
January 2006
Roche, A.M.
c14873d3-46ac-4df0-b178-b21e4835c8c4
James, M.F.M.
55614d48-1b93-4bbc-ac45-ccb28e0c384f
Grocott, M.P.W.
1e87b741-513e-4a22-be13-0f7bb344e8c2
Mythen, M.G.
c82ab585-86d5-44f8-bd3c-ead87c59d459
Roche, A.M., James, M.F.M., Grocott, M.P.W. and Mythen, M.G.
(2006)
Just scratching the surface: varied coagulation effects of polymer containers on TEG variables.
European Journal of Anaesthesiology, 23 (1), .
(PMID:16390565)
Abstract
Background and objective: Different types of polymer surfaces affect the activation of platelets and coagulation pathway containers depending on their surface qualities. Importantly, this could produce variability of coagulation results obtained with thrombelastographical analysis. We assessed the effects of blood storage on thrombelastograph, TEG, variables using polypropylene and polycarbonate containers.
Methods: An in vitro experiment was performed, with eight volunteers in each limb. Fresh whole blood was stored in polypropylene or polycarbonate tubes prior to TEGanalysis, to assess the role of these plastics in the TEG results obtained.
Results: The polycarbonate tubes displayed slower onset of coagulation and greater variability of data for all four basic TEG variables (r-time, k-time, alpha-angle and maximum amplitude, P < 0.05). Polycarbonate results fell outside manufacturer reference ranges.
Conclusions: It is likely that this is due to the altered surface properties and charge effects of the containers affecting proteins and platelets differently. Caution should be used in choosing which containers are used for storage of fresh blood prior to coagulation assessment, as variable results will follow where different types of plastic containers are employed.
This record has no associated files available for download.
More information
Published date: January 2006
Organisations:
Human Development & Health
Identifiers
Local EPrints ID: 348950
URI: http://eprints.soton.ac.uk/id/eprint/348950
ISSN: 0265-0215
PURE UUID: b99e2c5c-5f2c-4a2b-8806-a35a240db803
Catalogue record
Date deposited: 04 Mar 2013 15:16
Last modified: 08 Jan 2022 03:09
Export record
Contributors
Author:
A.M. Roche
Author:
M.F.M. James
Author:
M.G. Mythen
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics