Studies of the cavitational effects of clinical ultrasound by sonoluminescence: 1. Correlation of sonoluminescence with the standing wave pattern in an acoustic field produced by a therapeutic unit
Studies of the cavitational effects of clinical ultrasound by sonoluminescence: 1. Correlation of sonoluminescence with the standing wave pattern in an acoustic field produced by a therapeutic unit
A Therasonic 1030 (Electro-Medical Supplies) therapeutic ultrasound generator operating at 1 MHz continuous wave was used to insonate aerated water at two temperatures, 22 C and 37 C. Using various acoustically reflecting materials, sound fields were set up with different standing wave components. Measurements of the acoustic pressure variations on the axis of the sound fields were made using a needle hydrophone and the results were compared with photographs of the spatial distributions of the image intensified sonoluminescent light output. The near field region was used, thereby simulating the clinical situation.
Sustained sonoluminescence was observed for nominal intensities of 3 W cm-2, and acoustic reflections of greater than 40%. Under these conditions, if sonoluminescence did not appear spontaneously it could always be induced by rotating the transducer. Whenever bands of maximum light output formed they correlated closely with the pressure antinodes in the standing wave pattern. Very little light was produced by travelling wave fields.
1239-1248
Leighton, T.G.
3e5262ce-1d7d-42eb-b013-fcc5c286bbae
Pickworth, M.J.W.
8a0e0f89-fb53-4538-abb8-507923eeaf08
Walton, A.J.
2ac47979-cc42-4acd-9828-ead8a7897e85
Dendy, P.P.
ee1144be-8597-4abb-b01d-54470a991ba1
1988
Leighton, T.G.
3e5262ce-1d7d-42eb-b013-fcc5c286bbae
Pickworth, M.J.W.
8a0e0f89-fb53-4538-abb8-507923eeaf08
Walton, A.J.
2ac47979-cc42-4acd-9828-ead8a7897e85
Dendy, P.P.
ee1144be-8597-4abb-b01d-54470a991ba1
Leighton, T.G., Pickworth, M.J.W., Walton, A.J. and Dendy, P.P.
(1988)
Studies of the cavitational effects of clinical ultrasound by sonoluminescence: 1. Correlation of sonoluminescence with the standing wave pattern in an acoustic field produced by a therapeutic unit.
Physics in Medicine and Biology, 33 (11), .
(doi:10.1088/0031-9155/33/11/002).
Abstract
A Therasonic 1030 (Electro-Medical Supplies) therapeutic ultrasound generator operating at 1 MHz continuous wave was used to insonate aerated water at two temperatures, 22 C and 37 C. Using various acoustically reflecting materials, sound fields were set up with different standing wave components. Measurements of the acoustic pressure variations on the axis of the sound fields were made using a needle hydrophone and the results were compared with photographs of the spatial distributions of the image intensified sonoluminescent light output. The near field region was used, thereby simulating the clinical situation.
Sustained sonoluminescence was observed for nominal intensities of 3 W cm-2, and acoustic reflections of greater than 40%. Under these conditions, if sonoluminescence did not appear spontaneously it could always be induced by rotating the transducer. Whenever bands of maximum light output formed they correlated closely with the pressure antinodes in the standing wave pattern. Very little light was produced by travelling wave fields.
Text
# 1988 Leighton Pickworth Part 1 (PMB).pdf
- Version of Record
Restricted to Repository staff only
Request a copy
More information
Published date: 1988
Organisations:
Acoustics Group
Identifiers
Local EPrints ID: 349437
URI: http://eprints.soton.ac.uk/id/eprint/349437
PURE UUID: b37042b3-5610-44ca-bad0-bdec77fb857d
Catalogue record
Date deposited: 05 Mar 2013 14:06
Last modified: 15 Mar 2024 02:45
Export record
Altmetrics
Contributors
Author:
M.J.W. Pickworth
Author:
A.J. Walton
Author:
P.P. Dendy
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics