Parameters representing muscle tone, elasticity and stiffness of biceps brachii in healthy older males: symmetry and within-session reliability using the MyotonPRO

Parameters representing muscle tone, elasticity and stiffness of biceps brachii in healthy older males: symmetry and within-session reliability using the MyotonPRO
Parameters representing muscle tone, elasticity and stiffness of biceps brachii in healthy older males: symmetry and within-session reliability using the MyotonPRO
Background: Clinical assessments of muscle tone are subjective, often using the non-affected side for comparison. The MyotonPRO offers portable, non-invasive, objective measurement of mechanical properties of muscles.

Objective: This study aimed to investigate between-limb symmetry for mechanical properties of biceps brachii (BB) in older males and within-session intra-rater reliability of a novice user of the MyotonPRO device.

Methods: Twenty community-dwelling, right-handed males aged 65-85 years (mean 71.7; SD ± 4.9) were studied.

Exclusion criteria: history of neurological and musculoskeletal conditions, medications affecting muscle tone, and body mass index >30 kg/m2. The device applied 10 mechanical impulses at one second intervals, producing damped oscillations, from which frequency (non-neural tone), stiffness and logarithmic decrement (elasticity) were measured. With the participant resting supine, two consecutive sets of 10 recordings were taken bilaterally from BB.

Results: Percentage mean differences between-sides for larger and smaller values were 12% (tone), 14% (stiffness) and 27% (elasticity), which were statistically significant (p<0.001). Within-session reliability was excellent for all three parameters (ICC 3,2: all 0.99). Bland and Altman plots confirmed good agreement, without bias.

Conclusion: Symmetry of BB mechanical properties in a group of older males was less than 15% for tone and stiffness but not elasticity. Within-session intra-rater reliability of a novice user was excellent for all three parameters. Reliability over different days needs to be investigated. These findings indicate potential clinical application of the MyotonPRO for assessing abnormalities of muscle parameters in patients with neurological or musculoskeletal conditions, using comparison with the contralateral side (absolute difference), as well as databases of normative reference values from healthy control groups of different ages, gender and activity levels.
1-7
Bailey, L.
b67d7716-d1ff-4ceb-bcac-4038ff8000ca
Samuel, D.
03b00738-9b9c-4c0a-a85a-cf43fc0932fc
Warner, M.B.
f4dce73d-fb87-4f71-a3f0-078123aa040c
Stokes, M.
71730503-70ce-4e67-b7ea-a3e54579717f
Bailey, L.
b67d7716-d1ff-4ceb-bcac-4038ff8000ca
Samuel, D.
03b00738-9b9c-4c0a-a85a-cf43fc0932fc
Warner, M.B.
f4dce73d-fb87-4f71-a3f0-078123aa040c
Stokes, M.
71730503-70ce-4e67-b7ea-a3e54579717f

Bailey, L., Samuel, D., Warner, M.B. and Stokes, M. (2013) Parameters representing muscle tone, elasticity and stiffness of biceps brachii in healthy older males: symmetry and within-session reliability using the MyotonPRO. Journal of Neurological Disorders, 1 (1), 1-7. (doi:10.4172/jnd.1000116).

Record type: Article

Abstract

Background: Clinical assessments of muscle tone are subjective, often using the non-affected side for comparison. The MyotonPRO offers portable, non-invasive, objective measurement of mechanical properties of muscles.

Objective: This study aimed to investigate between-limb symmetry for mechanical properties of biceps brachii (BB) in older males and within-session intra-rater reliability of a novice user of the MyotonPRO device.

Methods: Twenty community-dwelling, right-handed males aged 65-85 years (mean 71.7; SD ± 4.9) were studied.

Exclusion criteria: history of neurological and musculoskeletal conditions, medications affecting muscle tone, and body mass index >30 kg/m2. The device applied 10 mechanical impulses at one second intervals, producing damped oscillations, from which frequency (non-neural tone), stiffness and logarithmic decrement (elasticity) were measured. With the participant resting supine, two consecutive sets of 10 recordings were taken bilaterally from BB.

Results: Percentage mean differences between-sides for larger and smaller values were 12% (tone), 14% (stiffness) and 27% (elasticity), which were statistically significant (p<0.001). Within-session reliability was excellent for all three parameters (ICC 3,2: all 0.99). Bland and Altman plots confirmed good agreement, without bias.

Conclusion: Symmetry of BB mechanical properties in a group of older males was less than 15% for tone and stiffness but not elasticity. Within-session intra-rater reliability of a novice user was excellent for all three parameters. Reliability over different days needs to be investigated. These findings indicate potential clinical application of the MyotonPRO for assessing abnormalities of muscle parameters in patients with neurological or musculoskeletal conditions, using comparison with the contralateral side (absolute difference), as well as databases of normative reference values from healthy control groups of different ages, gender and activity levels.

Text
__soton.ac.uk_ude_personalfiles_users_ds7_mydesktop_Bailey et al_JND.pdf - Other
Download (1MB)

More information

Published date: 15 April 2013
Organisations: Faculty of Health Sciences

Identifiers

Local EPrints ID: 352058
URI: http://eprints.soton.ac.uk/id/eprint/352058
PURE UUID: f7cc21b8-b847-44b4-bce0-61c363c310cb
ORCID for D. Samuel: ORCID iD orcid.org/0000-0003-3610-8032
ORCID for M.B. Warner: ORCID iD orcid.org/0000-0002-1483-0561
ORCID for M. Stokes: ORCID iD orcid.org/0000-0002-4204-0890

Catalogue record

Date deposited: 01 May 2013 13:59
Last modified: 17 Apr 2024 01:39

Export record

Altmetrics

Contributors

Author: L. Bailey
Author: D. Samuel ORCID iD
Author: M.B. Warner ORCID iD
Author: M. Stokes ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×