Measuring the adhesion of Ti/HA coatings to non-metallic implant materials
Measuring the adhesion of Ti/HA coatings to non-metallic implant materials
Introduction: Novel biomaterials may offer alternatives to metal arthroplasty bearings. To employ these materials in thin, bone conserving implants would require direct fixation to bone, using Titanium/HA coatings. Standard tests are used to evaluate the adhesion strength of coatings to metal substrates [1], versus FDA pass criteria [2]. In tensile adhesion testing, a disc is coated and uniform, uniaxial tension is exerted upon the coating-substrate interface; the strength is calculated from the failure load and surface area. Rapid failure occurs when the peak interface stress exceeds the adhesion strength, as local failure will propagate into an increasing tensile stress field.
Ceramics and reinforced polymers (e.g. carbon-fibre-reinforced PEEK), have considerably different stiffness (E) and Poisson's Ratio (?) from the coating and implant metals. We hypothesised that this substrate-coating stiffness mismatch would produce stress concentrations at the interface edge, well in excess of the uniform stress experienced with coatings on similar stiffness metals.
Methodology: The interface tensile stress field was predicted for the ASTM F1147 tensile strength test with a finite element analysis model, with a 500 ?m thick coating (50 ?m dense Ti layer, 450 ?m porous Ti/HA/adhesive layer), bonded to a stainless steel headpiece with FM1000 adhesive (Fig. 1). Solutions were obtained for:
Configuration A: ASTM-standard geometry with Ti-6Al-4V (E = 110GPa, ? = 0.31), CoCrMo (E = 196GPa, ? = 0.30), ceramic (E = 350GPa, ? = 0.22, e.g. BIOLOX delta) and CFR-PEEK (E = 15GPa, ? = 0.41, e.g. Invibio MOTIS) substrates.
Modified models were used to analyse oversized substrate discs:
Configuration B: coated fully and bonded to the standard diameter headpiece, and
Configuration C: Coated only where bonded to the headpiece.
Results and Discussion: The stiffness mismatch between the coating and the ceramic and CFR-PEEK substrates was predicted to introduce, respectively, a 1.80x and 3.57x stress concentration compared to a Ti6Al4V substrate (Fig. 2), thereby reducing the failure load for a given interface strength. These predictions consider the test stress distribution only, and do not assess the coating-substrate interface strength. However, the failure load is a function of the interface strength and the peak test stress, so the standard test and stress calculation for stiffness-mismatched substrates may indicate artificially low adhesion strength.
The test may be modified to suit a particular material combination. As an example, for ceramic substrates the results indicate that an oversized, fully coated specimen (B) would experience stress closest to the standard's intended uniform stress field, suggesting that this configuration would be more appropriate. The stress distribution may be sensitive to the coating thickness, so tests should be verified accordingly.
Conclusion: The ASTM coating tensile adhesion strength test standard was predicted to generate a non-uniform interfacial stress for ceramic and polymer composite substrate materials. The standard may not be directly applicable for non-metal substrates as the stiffness mismatch needs to be considered.
286
Dickinson, Alexander
10151972-c1b5-4f7d-bc12-6482b5870cad
Taylor, Andrew
39974814-4868-4c73-a3fa-2adfa4be3e46
Roques, Anne
5560237b-df68-492e-a4a9-0a075eaf104b
Browne, Martin
6578cc37-7bd6-43b9-ae5c-77ccb7726397
31 December 2013
Dickinson, Alexander
10151972-c1b5-4f7d-bc12-6482b5870cad
Taylor, Andrew
39974814-4868-4c73-a3fa-2adfa4be3e46
Roques, Anne
5560237b-df68-492e-a4a9-0a075eaf104b
Browne, Martin
6578cc37-7bd6-43b9-ae5c-77ccb7726397
Dickinson, Alexander, Taylor, Andrew, Roques, Anne and Browne, Martin
(2013)
Measuring the adhesion of Ti/HA coatings to non-metallic implant materials.
[in special issue: International Society for Technology in Arthroplasty (ISTA) Palm Beach, Florida, USA, 16–19 October 2013]
Orthopaedic Proceedings, 95-B, supplement 34, .
Abstract
Introduction: Novel biomaterials may offer alternatives to metal arthroplasty bearings. To employ these materials in thin, bone conserving implants would require direct fixation to bone, using Titanium/HA coatings. Standard tests are used to evaluate the adhesion strength of coatings to metal substrates [1], versus FDA pass criteria [2]. In tensile adhesion testing, a disc is coated and uniform, uniaxial tension is exerted upon the coating-substrate interface; the strength is calculated from the failure load and surface area. Rapid failure occurs when the peak interface stress exceeds the adhesion strength, as local failure will propagate into an increasing tensile stress field.
Ceramics and reinforced polymers (e.g. carbon-fibre-reinforced PEEK), have considerably different stiffness (E) and Poisson's Ratio (?) from the coating and implant metals. We hypothesised that this substrate-coating stiffness mismatch would produce stress concentrations at the interface edge, well in excess of the uniform stress experienced with coatings on similar stiffness metals.
Methodology: The interface tensile stress field was predicted for the ASTM F1147 tensile strength test with a finite element analysis model, with a 500 ?m thick coating (50 ?m dense Ti layer, 450 ?m porous Ti/HA/adhesive layer), bonded to a stainless steel headpiece with FM1000 adhesive (Fig. 1). Solutions were obtained for:
Configuration A: ASTM-standard geometry with Ti-6Al-4V (E = 110GPa, ? = 0.31), CoCrMo (E = 196GPa, ? = 0.30), ceramic (E = 350GPa, ? = 0.22, e.g. BIOLOX delta) and CFR-PEEK (E = 15GPa, ? = 0.41, e.g. Invibio MOTIS) substrates.
Modified models were used to analyse oversized substrate discs:
Configuration B: coated fully and bonded to the standard diameter headpiece, and
Configuration C: Coated only where bonded to the headpiece.
Results and Discussion: The stiffness mismatch between the coating and the ceramic and CFR-PEEK substrates was predicted to introduce, respectively, a 1.80x and 3.57x stress concentration compared to a Ti6Al4V substrate (Fig. 2), thereby reducing the failure load for a given interface strength. These predictions consider the test stress distribution only, and do not assess the coating-substrate interface strength. However, the failure load is a function of the interface strength and the peak test stress, so the standard test and stress calculation for stiffness-mismatched substrates may indicate artificially low adhesion strength.
The test may be modified to suit a particular material combination. As an example, for ceramic substrates the results indicate that an oversized, fully coated specimen (B) would experience stress closest to the standard's intended uniform stress field, suggesting that this configuration would be more appropriate. The stress distribution may be sensitive to the coating thickness, so tests should be verified accordingly.
Conclusion: The ASTM coating tensile adhesion strength test standard was predicted to generate a non-uniform interfacial stress for ceramic and polymer composite substrate materials. The standard may not be directly applicable for non-metal substrates as the stiffness mismatch needs to be considered.
Text
Dickinson ISTA.pdf
- Other
More information
Published date: 31 December 2013
Organisations:
Bioengineering Group
Identifiers
Local EPrints ID: 360840
URI: http://eprints.soton.ac.uk/id/eprint/360840
ISSN: 1358-992X
PURE UUID: df6255b6-64a7-408d-af0b-77fdb5b2ceba
Catalogue record
Date deposited: 08 Jan 2014 10:13
Last modified: 15 Mar 2024 03:27
Export record
Contributors
Author:
Andrew Taylor
Author:
Anne Roques
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics