Intermittent pneumatic compression. A comparison of femoral vein velocity with five different devices
Intermittent pneumatic compression. A comparison of femoral vein velocity with five different devices
Aim: different mechanical devices for tbromboprophylaxis have different flow characteristics. A new device (Vadoplex) has been developed to provide a short impulse around the calf, a concept derived from the efficacy of foot impulse technology. New devices should be compared with existing devices to establish whether it has a comparable ability to augment venous velocity. Objectives of the study were to compare the venous velocity induced by the Vadoplex with established intermittent pneumatic calf and leg compressors (Covidien and Huntleigh).
Methods: the venous velocity was established in ten healthy volunteers with standardised Duplex ultraonography of the common femoral vein. Measurements were taken at rest and on standing, with each device inactive and active.
Results: the Vadoplex induced an increased femoral venous velocity at least as enhanced as established calf and full leg sleeves.
Conclusion: calf impulse technology is an alternative to other systems in enhancing femoral vein blood flow, itself a surrogate for assumed thromboprophylactic effect.
404-409
Warwick, David J.
d23cacce-41eb-42bd-b8cc-da6a3b837a9f
Dewbury, K.
d51f5c0d-5d22-4993-aa08-cba264e9dabf
Forrester, A.I.J.
176bf191-3fc2-46b4-80e0-9d9a0cd7a572
August 2013
Warwick, David J.
d23cacce-41eb-42bd-b8cc-da6a3b837a9f
Dewbury, K.
d51f5c0d-5d22-4993-aa08-cba264e9dabf
Forrester, A.I.J.
176bf191-3fc2-46b4-80e0-9d9a0cd7a572
Warwick, David J., Dewbury, K. and Forrester, A.I.J.
(2013)
Intermittent pneumatic compression. A comparison of femoral vein velocity with five different devices.
International Angiology, 32 (4), .
(PMID:23822943)
Abstract
Aim: different mechanical devices for tbromboprophylaxis have different flow characteristics. A new device (Vadoplex) has been developed to provide a short impulse around the calf, a concept derived from the efficacy of foot impulse technology. New devices should be compared with existing devices to establish whether it has a comparable ability to augment venous velocity. Objectives of the study were to compare the venous velocity induced by the Vadoplex with established intermittent pneumatic calf and leg compressors (Covidien and Huntleigh).
Methods: the venous velocity was established in ten healthy volunteers with standardised Duplex ultraonography of the common femoral vein. Measurements were taken at rest and on standing, with each device inactive and active.
Results: the Vadoplex induced an increased femoral venous velocity at least as enhanced as established calf and full leg sleeves.
Conclusion: calf impulse technology is an alternative to other systems in enhancing femoral vein blood flow, itself a surrogate for assumed thromboprophylactic effect.
This record has no associated files available for download.
More information
Published date: August 2013
Organisations:
Faculty of Medicine
Identifiers
Local EPrints ID: 364896
URI: http://eprints.soton.ac.uk/id/eprint/364896
ISSN: 0392-9590
PURE UUID: ff5abb68-aea7-47c0-8d0a-fdfef3b3b1a4
Catalogue record
Date deposited: 16 May 2014 08:14
Last modified: 08 Jan 2022 18:10
Export record
Contributors
Author:
David J. Warwick
Author:
K. Dewbury
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.