Modelling the effects of cochlear implant current focusing

Modelling the effects of cochlear implant current focusing
Modelling the effects of cochlear implant current focusing
Objectives
A finite element model of the human coiled cochlea was used to model the voltage distribution due to stimulation by the individual electrodes of a cochlear implant. The scalar position of the electrode array was also varied in order to investigate its effect on the voltage distribution. Multi-electrode current focusing methods were then investigated, with the aim of increasing spatial selectivity.

Methods
Simultaneous current focusing is initially achieved, as in previous publications, by calculating the input currents to the 22 electrodes that best separates the voltages at these electrode positions. The benefits of this electrode focusing strategy do not, however, entirely carry over to the predicted voltage distributions at the position of the spiral ganglion cells, where excitation is believed to occur. A novel focusing strategy is then simulated, which compensates for the impedances between the currents at the electrode sites and the voltage distribution directly at the position of the spiral ganglion cells.


Results
The new strategy produces much better focusing at the sites of the spiral ganglion cells, as expected, but at the cost of increased current requirements. Regularization was introduced in order to reduce current requirements, which also reduced the sensitivity of the solution to uncertainties in the impedance matrix, so that improved focusing was achieved with similar current requirements to that of electrode focusing.


Discussion
Although such focusing strategies cannot be achieved in practice at the moment, since the responses from the electrodes to the neural sites cannot be determined with currently available recording methods, these results do support the feasibility of a more effective focusing strategy, which may provide improved spectral resolution leading to improved perception of sound.
cochlear implant modelling, stimulation focusing, coiled cochlea, finite element analysis
1467-0100
318-326
Saba, Rami
ec4ed189-d390-47c4-aa52-86401b61f27e
Elliott, Stephen J.
721dc55c-8c3e-4895-b9c4-82f62abd3567
Wang, Shouyan
fa12f1bf-cac9-4118-abdd-9d52f235b05c
Saba, Rami
ec4ed189-d390-47c4-aa52-86401b61f27e
Elliott, Stephen J.
721dc55c-8c3e-4895-b9c4-82f62abd3567
Wang, Shouyan
fa12f1bf-cac9-4118-abdd-9d52f235b05c

Saba, Rami, Elliott, Stephen J. and Wang, Shouyan (2014) Modelling the effects of cochlear implant current focusing. Cochlear Implants International, 15 (6), 318-326. (doi:10.1179/1754762814Y.0000000081).

Record type: Article

Abstract

Objectives
A finite element model of the human coiled cochlea was used to model the voltage distribution due to stimulation by the individual electrodes of a cochlear implant. The scalar position of the electrode array was also varied in order to investigate its effect on the voltage distribution. Multi-electrode current focusing methods were then investigated, with the aim of increasing spatial selectivity.

Methods
Simultaneous current focusing is initially achieved, as in previous publications, by calculating the input currents to the 22 electrodes that best separates the voltages at these electrode positions. The benefits of this electrode focusing strategy do not, however, entirely carry over to the predicted voltage distributions at the position of the spiral ganglion cells, where excitation is believed to occur. A novel focusing strategy is then simulated, which compensates for the impedances between the currents at the electrode sites and the voltage distribution directly at the position of the spiral ganglion cells.


Results
The new strategy produces much better focusing at the sites of the spiral ganglion cells, as expected, but at the cost of increased current requirements. Regularization was introduced in order to reduce current requirements, which also reduced the sensitivity of the solution to uncertainties in the impedance matrix, so that improved focusing was achieved with similar current requirements to that of electrode focusing.


Discussion
Although such focusing strategies cannot be achieved in practice at the moment, since the responses from the electrodes to the neural sites cannot be determined with currently available recording methods, these results do support the feasibility of a more effective focusing strategy, which may provide improved spectral resolution leading to improved perception of sound.

This record has no associated files available for download.

More information

Published date: November 2014
Keywords: cochlear implant modelling, stimulation focusing, coiled cochlea, finite element analysis
Organisations: Signal Processing & Control Grp

Identifiers

Local EPrints ID: 370697
URI: http://eprints.soton.ac.uk/id/eprint/370697
ISSN: 1467-0100
PURE UUID: 96bb4cac-8dc2-43a3-bc6d-b0d851926098

Catalogue record

Date deposited: 03 Nov 2014 16:33
Last modified: 14 Mar 2024 18:20

Export record

Altmetrics

Contributors

Author: Rami Saba
Author: Shouyan Wang

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×