Hands-on defibrillation: theoretical and practical aspects of patient and rescuer safety

Hands-on defibrillation: theoretical and practical aspects of patient and rescuer safety
Hands-on defibrillation: theoretical and practical aspects of patient and rescuer safety
Defibrillators are used to treat many thousands of people each year using very high voltages, but, despite this, reported injuries to rescuers are rare. Although even a small number of reported injuries is not ideal, the safety record of the defibrillator using the current protocol is widely regarded as being acceptable.

There is increasing evidence that clinical outcome is significantly improved with continuous chest compressions, but defibrillation is a common cause of interruptions; even short interruptions, such as those associated with defibrillation, may detrimentally affect the outcome. This has led to discussions regarding the possibility of continuing chest compressions during defibrillation; a process involving a rescuer working in close proximity to voltages of up to 5000 V.

Not only do voltages of this magnitude have significant implications for the rescuer performing chest compressions, but there are also risks to other rescuers in the proximity, the patient and other bystanders. Clearly any deviation from accepted practice should only be undertaken following careful consideration of the risks and benefits to the patient, rescuers and others.

This review summarises the physical principles of electrical risk and identifies ways in which these could be managed. In doing so, it is hoped that in future it may be possible to deliver continuous and safe manual chest compressions during defibrillator discharge in order to improve patient outcome.
defibrillation, safety
0300-9572
551-556
Petley, Graham W.
4f2da40b-3c7b-4adc-b75c-e24e62bb1cf0
Cotton, Angela M.
eed9d254-747c-43f9-99f5-fee86afc918f
Deakin, Charles D.
560d993b-bbc9-4548-9990-272ed18a011d
Petley, Graham W.
4f2da40b-3c7b-4adc-b75c-e24e62bb1cf0
Cotton, Angela M.
eed9d254-747c-43f9-99f5-fee86afc918f
Deakin, Charles D.
560d993b-bbc9-4548-9990-272ed18a011d

Petley, Graham W., Cotton, Angela M. and Deakin, Charles D. (2012) Hands-on defibrillation: theoretical and practical aspects of patient and rescuer safety. Resuscitation, 83 (5), 551-556. (doi:10.1016/j.resuscitation.2011.11.005). (PMID:22094984)

Record type: Article

Abstract

Defibrillators are used to treat many thousands of people each year using very high voltages, but, despite this, reported injuries to rescuers are rare. Although even a small number of reported injuries is not ideal, the safety record of the defibrillator using the current protocol is widely regarded as being acceptable.

There is increasing evidence that clinical outcome is significantly improved with continuous chest compressions, but defibrillation is a common cause of interruptions; even short interruptions, such as those associated with defibrillation, may detrimentally affect the outcome. This has led to discussions regarding the possibility of continuing chest compressions during defibrillation; a process involving a rescuer working in close proximity to voltages of up to 5000 V.

Not only do voltages of this magnitude have significant implications for the rescuer performing chest compressions, but there are also risks to other rescuers in the proximity, the patient and other bystanders. Clearly any deviation from accepted practice should only be undertaken following careful consideration of the risks and benefits to the patient, rescuers and others.

This review summarises the physical principles of electrical risk and identifies ways in which these could be managed. In doing so, it is hoped that in future it may be possible to deliver continuous and safe manual chest compressions during defibrillator discharge in order to improve patient outcome.

This record has no associated files available for download.

More information

Published date: May 2012
Keywords: defibrillation, safety
Organisations: Physics & Astronomy, Clinical & Experimental Sciences

Identifiers

Local EPrints ID: 383362
URI: http://eprints.soton.ac.uk/id/eprint/383362
ISSN: 0300-9572
PURE UUID: c69504e2-95cf-46ff-979a-188cb7d6ffd9
ORCID for Graham W. Petley: ORCID iD orcid.org/0000-0002-3295-0444

Catalogue record

Date deposited: 28 Oct 2015 14:39
Last modified: 15 Mar 2024 02:53

Export record

Altmetrics

Contributors

Author: Angela M. Cotton

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×