Interfacial pressure and shear sensor systems for lower limb prosthetic applications

Interfacial pressure and shear sensor systems for lower limb prosthetic applications
Interfacial pressure and shear sensor systems for lower limb prosthetic applications
Lower limb prosthetic socket provides the interface for the transfer of loads between the ground and the stump. During daily physical activities, the stump has to endure mechanical stresses in normal (pressure) and tangential (shear) directions to the stump/socket interface, both of which may cause discomfort, pain, and a host of stump health problems. Despite the importance of the issues, there is currently no clinically-friendly sensor technology available to monitor both pressure and shear at the stump/socket interface.

In this thesis, two sensor systems were developed - both capable of simultaneously measuring dynamic pressure and shear at the lower limb stump/socket interface. In particular, two thin, capacitive sensors based on flexible elastomers (3D-printed and silicone) were designed and fabricated. To acquire sensor signals, appropriate data acquisition circuitries and PC software were produced. A viscoelastic model for real-time processing of the sensor signals was developed, to improve their dynamic response. The sensor systems were characterised in research lab settings, and their performance was verified against comprehensive design requirements.

The characterised sensor systems were validated at the stump/socket interface in a series of amputee walking tests, and a range of clinical factors were investigated. Pressure and shear were measured at up to three discrete locations of a trans-femoral stump. Their temporal profiles when walking on level and inclined surfaces were obtained. The results suggest that ankle flexion resistance may have an effect on both the magnitudes and the temporal profiles of pressure and shear at the stump/socket interface.

The thesis delivers a tool for measuring pressure and shear at the stump/socket interface in a clinically-friendly manner. What is more, it contributes to the prosthetic field by bringing an insight into the pressure and shear at the trans-femoral stump/socket interface, thus enhancing the understanding of its biomechanics. In the future such a sensor technology could potentially aid daily monitoring of socket fit, assist prosthetists with socket fitting, support research on the stump/socket interface biomechanics, etc.
Laszczak, Piotr
ea4976f2-006b-4be6-b49a-9d4a88c1a4bf
Laszczak, Piotr
ea4976f2-006b-4be6-b49a-9d4a88c1a4bf
Jiang, Liudi
374f2414-51f0-418f-a316-e7db0d6dc4d1

Laszczak, Piotr (2016) Interfacial pressure and shear sensor systems for lower limb prosthetic applications. University of Southampton, Faculty of Engineering and the Environment, Doctoral Thesis, 400pp.

Record type: Thesis (Doctoral)

Abstract

Lower limb prosthetic socket provides the interface for the transfer of loads between the ground and the stump. During daily physical activities, the stump has to endure mechanical stresses in normal (pressure) and tangential (shear) directions to the stump/socket interface, both of which may cause discomfort, pain, and a host of stump health problems. Despite the importance of the issues, there is currently no clinically-friendly sensor technology available to monitor both pressure and shear at the stump/socket interface.

In this thesis, two sensor systems were developed - both capable of simultaneously measuring dynamic pressure and shear at the lower limb stump/socket interface. In particular, two thin, capacitive sensors based on flexible elastomers (3D-printed and silicone) were designed and fabricated. To acquire sensor signals, appropriate data acquisition circuitries and PC software were produced. A viscoelastic model for real-time processing of the sensor signals was developed, to improve their dynamic response. The sensor systems were characterised in research lab settings, and their performance was verified against comprehensive design requirements.

The characterised sensor systems were validated at the stump/socket interface in a series of amputee walking tests, and a range of clinical factors were investigated. Pressure and shear were measured at up to three discrete locations of a trans-femoral stump. Their temporal profiles when walking on level and inclined surfaces were obtained. The results suggest that ankle flexion resistance may have an effect on both the magnitudes and the temporal profiles of pressure and shear at the stump/socket interface.

The thesis delivers a tool for measuring pressure and shear at the stump/socket interface in a clinically-friendly manner. What is more, it contributes to the prosthetic field by bringing an insight into the pressure and shear at the trans-femoral stump/socket interface, thus enhancing the understanding of its biomechanics. In the future such a sensor technology could potentially aid daily monitoring of socket fit, assist prosthetists with socket fitting, support research on the stump/socket interface biomechanics, etc.

Text
Final E-Thesis for E-Prints.pdf - Other
Restricted to Repository staff only until 31 May 2025.
Available under License University of Southampton Thesis Licence.

More information

Published date: April 2016
Organisations: University of Southampton, Engineering Mats & Surface Engineerg Gp

Identifiers

Local EPrints ID: 393746
URI: http://eprints.soton.ac.uk/id/eprint/393746
PURE UUID: 8eef2d2e-552b-4d62-bf07-bfbd44dc9e21
ORCID for Liudi Jiang: ORCID iD orcid.org/0000-0002-3400-825X

Catalogue record

Date deposited: 06 Jul 2016 10:39
Last modified: 15 Mar 2024 03:24

Export record

Contributors

Author: Piotr Laszczak
Thesis advisor: Liudi Jiang ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×