Validation of a vibrating-wire viscometer: measurements in the range of 0.5 to 135 mPa·s

Validation of a vibrating-wire viscometer: measurements in the range of 0.5 to 135 mPa·s
Validation of a vibrating-wire viscometer: measurements in the range of 0.5 to 135 mPa·s
The present article describes a novel vibrating-wire viscometer that has been shown to be able to measure viscosities up to 135 mPa·s, after calibration against water at 20 C, to provide direct traceability to the primary reference for viscosity. For the purpose of validating the instrument, measurements of the viscosity of some selected fluids, including standard reference liquids with viscosity on the order of 100 mPa·s at 20 C and 2,2,4-trimethylpentane with a viscosity of about 0.5 mPa·s at the same temperature, were performed. The results obtained show that the instrument is capable of performing viscosity measurements with an estimated overall uncertainty better than ±0.8% over the range of (0.5 to 135) mPa·s
0021-9568
201-205
Caetano, Fernando J.P.
cbcae1ed-2f47-4e49-8169-f996596fba71
Fareleira, João M.N.A.
82d525d0-0828-4af7-917b-b92c986945c2
Oliveira, Carla M.B.P.
2eaaaa28-0dfd-4a9b-9b0f-cbe7fced80b5
Wakeham, William A.
88549729-a39a-497f-b112-feaa6be2c449
Caetano, Fernando J.P.
cbcae1ed-2f47-4e49-8169-f996596fba71
Fareleira, João M.N.A.
82d525d0-0828-4af7-917b-b92c986945c2
Oliveira, Carla M.B.P.
2eaaaa28-0dfd-4a9b-9b0f-cbe7fced80b5
Wakeham, William A.
88549729-a39a-497f-b112-feaa6be2c449

Caetano, Fernando J.P., Fareleira, João M.N.A., Oliveira, Carla M.B.P. and Wakeham, William A. (2005) Validation of a vibrating-wire viscometer: measurements in the range of 0.5 to 135 mPa·s. Journal of Chemical & Engineering Data, 50 (1), 201-205. (doi:10.1021/je049733f).

Record type: Article

Abstract

The present article describes a novel vibrating-wire viscometer that has been shown to be able to measure viscosities up to 135 mPa·s, after calibration against water at 20 C, to provide direct traceability to the primary reference for viscosity. For the purpose of validating the instrument, measurements of the viscosity of some selected fluids, including standard reference liquids with viscosity on the order of 100 mPa·s at 20 C and 2,2,4-trimethylpentane with a viscosity of about 0.5 mPa·s at the same temperature, were performed. The results obtained show that the instrument is capable of performing viscosity measurements with an estimated overall uncertainty better than ±0.8% over the range of (0.5 to 135) mPa·s

This record has no associated files available for download.

More information

Published date: 2005

Identifiers

Local EPrints ID: 39571
URI: http://eprints.soton.ac.uk/id/eprint/39571
ISSN: 0021-9568
PURE UUID: 9b70e5f4-d26a-4820-ad80-7fc605ba7957

Catalogue record

Date deposited: 28 Jun 2006
Last modified: 15 Mar 2024 08:15

Export record

Altmetrics

Contributors

Author: Fernando J.P. Caetano
Author: João M.N.A. Fareleira
Author: Carla M.B.P. Oliveira
Author: William A. Wakeham

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×