Evaluation of the Metasin assay for intraoperative assessment of sentinel lymph node metastases in breast cancer

Evaluation of the Metasin assay for intraoperative assessment of sentinel lymph node metastases in breast cancer
Evaluation of the Metasin assay for intraoperative assessment of sentinel lymph node metastases in breast cancer
Aims: Sentinel lymph node (SLN) biopsy is the preferred surgical technique for staging the axilla in clinically node-negative breast cancer. Accurate intraoperative staging allows for the immediate performance of an axillary clearance in node-positive patients. We assessed the Metasin assay for the intraoperative analysis of SLNs in a prospective evaluation of 250 consecutive patients undergoing intraoperative SLN analysis at the Breast Unit, University Hospital, Southampton, UK.

Methods: Metasin uses a quantitative reverse transcription PCR to detect two markers of metastasis: cytokeratin 19 (CK19) an epithelial marker and mammaglobin (MGB) a breast specific marker. Metasin results were compared with the results from routine paraffin block histopathology.

Results: Metasin was robust, with a failure rate of <1%, and demonstrated excellent accuracy and reproducibility. The average turnaround time for the Metasin assay was 42?min, the largest variable being the number of nodes assayed. A total of 533 SLNs were evaluated with 75 patients testing positive for MGB and/or CK19. Based on the analysis of individual SLNs, the overall concordance between Metasin and histology was 92.3% (sensitivity 88.7%, specificity 92.9%). When adjusted for tissue allocation bias, the concordance was 93.8% (sensitivity 89.8%, specificity 94.6%). In this evaluation, 57/250 patients (23%) proceeded to axillary clearance based on Metasin results and were considered spared a second operative procedure.

Conclusions: Metasin has proven to be an accurate, reproducible and reliable laboratory test. The analysis time is acceptable for intraoperative use, and in comparison to routine histology demonstrates acceptable concordance, sensitivity and specificity.
0021-9746
1-6
Smith, G.J.
fe8dcd9d-005d-44dc-bbb4-f380ccc6e544
Hodges, E.
85c114a2-4533-48d7-8a6c-7b82a4138251
Markham, H.
de7b2315-6701-4747-aa92-aa8cef4a551f
Zhang, S.
76640b2c-83c0-4a90-bcd8-939ffb2505aa
Cutress, R.I.
68ae4f86-e8cf-411f-a335-cdba51797406
Smith, G.J.
fe8dcd9d-005d-44dc-bbb4-f380ccc6e544
Hodges, E.
85c114a2-4533-48d7-8a6c-7b82a4138251
Markham, H.
de7b2315-6701-4747-aa92-aa8cef4a551f
Zhang, S.
76640b2c-83c0-4a90-bcd8-939ffb2505aa
Cutress, R.I.
68ae4f86-e8cf-411f-a335-cdba51797406

Smith, G.J., Hodges, E., Markham, H., Zhang, S. and Cutress, R.I. (2016) Evaluation of the Metasin assay for intraoperative assessment of sentinel lymph node metastases in breast cancer. Journal of Clinical Pathology, 1-6. (doi:10.1136/jclinpath-2016-203728). (PMID:27406050)

Record type: Article

Abstract

Aims: Sentinel lymph node (SLN) biopsy is the preferred surgical technique for staging the axilla in clinically node-negative breast cancer. Accurate intraoperative staging allows for the immediate performance of an axillary clearance in node-positive patients. We assessed the Metasin assay for the intraoperative analysis of SLNs in a prospective evaluation of 250 consecutive patients undergoing intraoperative SLN analysis at the Breast Unit, University Hospital, Southampton, UK.

Methods: Metasin uses a quantitative reverse transcription PCR to detect two markers of metastasis: cytokeratin 19 (CK19) an epithelial marker and mammaglobin (MGB) a breast specific marker. Metasin results were compared with the results from routine paraffin block histopathology.

Results: Metasin was robust, with a failure rate of <1%, and demonstrated excellent accuracy and reproducibility. The average turnaround time for the Metasin assay was 42?min, the largest variable being the number of nodes assayed. A total of 533 SLNs were evaluated with 75 patients testing positive for MGB and/or CK19. Based on the analysis of individual SLNs, the overall concordance between Metasin and histology was 92.3% (sensitivity 88.7%, specificity 92.9%). When adjusted for tissue allocation bias, the concordance was 93.8% (sensitivity 89.8%, specificity 94.6%). In this evaluation, 57/250 patients (23%) proceeded to axillary clearance based on Metasin results and were considered spared a second operative procedure.

Conclusions: Metasin has proven to be an accurate, reproducible and reliable laboratory test. The analysis time is acceptable for intraoperative use, and in comparison to routine histology demonstrates acceptable concordance, sensitivity and specificity.

Text
Evaluation of the Metasin assay for intraoperative assessment of sentinel lymph node metastases in breast cancer. 2016.docx - Accepted Manuscript
Download (443kB)

More information

Accepted/In Press date: 17 June 2016
e-pub ahead of print date: 12 July 2016
Organisations: Cancer Sciences

Identifiers

Local EPrints ID: 398151
URI: http://eprints.soton.ac.uk/id/eprint/398151
ISSN: 0021-9746
PURE UUID: 551db4fc-013b-4707-bcfe-2283d5eeef06

Catalogue record

Date deposited: 20 Jul 2016 10:37
Last modified: 15 Mar 2024 05:45

Export record

Altmetrics

Contributors

Author: G.J. Smith
Author: E. Hodges
Author: H. Markham
Author: S. Zhang
Author: R.I. Cutress

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×