An evaluation and comparison of cross-sensitivities of various different types of screen printed Ag/AgCl reference electrodes

An evaluation and comparison of cross-sensitivities of various different types of screen printed Ag/AgCl reference electrodes
An evaluation and comparison of cross-sensitivities of various different types of screen printed Ag/AgCl reference electrodes

Different types of thick film screen printed silver and silver-silver chloride (Ag/AgCl) electrodes have been evaluated for use as solid state alternatives to liquid electrolyte filled Ag/AgCl reference electrodes. The tested electrodes included those employing different types of binder (glass and polymer) for the Ag/AgCl layer, those consisting of a bare silver (pseudo reference) electrode and those with a bare silver electrode with an additional covering layer of a polymer bound KCl salt matrix. The cross-sensitivities of these different electrodes to several different ion species (e.g. chlorides, nitrates and phosphates) were examined by recording their electro potentials in varying ionic strength solutions.

cross sensitivity, ion selective electrodes, reference electrodes, thick film sensors
299-303
IEEE
Atkinson, John K.
5e9729b2-0e1f-400d-a889-c74f6390ea58
Atkinson, John K.
5e9729b2-0e1f-400d-a889-c74f6390ea58

Atkinson, John K. (2018) An evaluation and comparison of cross-sensitivities of various different types of screen printed Ag/AgCl reference electrodes. In 2018 International Conference on Electronics Packaging and iMAPS All Asia Conference (ICEP-IAAC). IEEE. pp. 299-303 . (doi:10.23919/ICEP.2018.8374308).

Record type: Conference or Workshop Item (Paper)

Abstract

Different types of thick film screen printed silver and silver-silver chloride (Ag/AgCl) electrodes have been evaluated for use as solid state alternatives to liquid electrolyte filled Ag/AgCl reference electrodes. The tested electrodes included those employing different types of binder (glass and polymer) for the Ag/AgCl layer, those consisting of a bare silver (pseudo reference) electrode and those with a bare silver electrode with an additional covering layer of a polymer bound KCl salt matrix. The cross-sensitivities of these different electrodes to several different ion species (e.g. chlorides, nitrates and phosphates) were examined by recording their electro potentials in varying ionic strength solutions.

This record has no associated files available for download.

More information

e-pub ahead of print date: 6 June 2018
Published date: 7 June 2018
Venue - Dates: 2018 International Conference on Electronics Packaging and iMAPS All Asia Conference (ICEP-IAAC), , Kuwana, Mie, Japan, 2018-04-17 - 2018-04-21
Keywords: cross sensitivity, ion selective electrodes, reference electrodes, thick film sensors

Identifiers

Local EPrints ID: 424824
URI: http://eprints.soton.ac.uk/id/eprint/424824
PURE UUID: 5310d189-12d6-4e82-adb1-ff7c633fe873
ORCID for John K. Atkinson: ORCID iD orcid.org/0000-0003-3411-8034

Catalogue record

Date deposited: 05 Oct 2018 11:48
Last modified: 16 Mar 2024 02:33

Export record

Altmetrics

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×