The University of Southampton
University of Southampton Institutional Repository

Photoacoustic and photothermal detection of trace compounds in water

Photoacoustic and photothermal detection of trace compounds in water
Photoacoustic and photothermal detection of trace compounds in water

The detection of trace compounds in water using photoacoustic or photothermal techniques is presented, with specific reference to on-line water quality monitoring. A key requirement was the use of simple, robust and inexpensive light sources, whose continuous-wave output was modulated electronically at relatively low frequencies (~100Hz). Two closed-cell, low-frequency optical absorption detectors have been developed and evaluated. In both cases, a novel theoretical treatment revealed that the optimum cell design had a pressure sensor whose mechanical compliance was equal to the bulk compliance of the enclosed water. High sensitivity was achieved despite the use of lower power light sources, potentially extending the applicability of photothermal/photoacoustic methods to low-cost systems.

The photoacoustic cell incorporated a thin-walled piezoelectric cylinder. Using a 678nm laser diode emitting 1.4±0.2 mW rms, a minimum detectable absorption coefficient of 3x10-3cm-1 was demonstrated.

Better performance was achieved using the photothermal detector. Photothermal expansion in aqueous samples caused the deflection of a water meniscus held across a 200 μm-radius pinhole. Meniscus displacement was monitored using fibre optic interferometry. The magnitude and form of photothermal meniscus deflection signals were shown to be in good agreement with theory. Absorption by 1.5ppb anthracene in water was detected using a mercury discharge lamp (254nm) as an excitation source, and light emitting diodes were used to excite photothermal signals in a liquid (water) for the first time. The technique was limited by light source emission intensities and environmental acoustic noise, for example giving a minimum detectable absorption coefficient of 2x10-4cm-1 using an LED emitting 0.38±0.06 mW rms.

The minimum detectable absorption coefficients of the photothermal detector and of a conventional transmission spectrometer were found to be similar, but the photothermal technique was far less sensitive to the effects of light scattering. The ability to detect trace contaminants in water, in the presence of turbidity, is of great importance to the water industry.

University of Southampton
Hodgkinson, Jane
ccb8c8c6-a374-4e1a-9997-83cc63531f58
Hodgkinson, Jane
ccb8c8c6-a374-4e1a-9997-83cc63531f58

Hodgkinson, Jane (1998) Photoacoustic and photothermal detection of trace compounds in water. University of Southampton, Doctoral Thesis.

Record type: Thesis (Doctoral)

Abstract

The detection of trace compounds in water using photoacoustic or photothermal techniques is presented, with specific reference to on-line water quality monitoring. A key requirement was the use of simple, robust and inexpensive light sources, whose continuous-wave output was modulated electronically at relatively low frequencies (~100Hz). Two closed-cell, low-frequency optical absorption detectors have been developed and evaluated. In both cases, a novel theoretical treatment revealed that the optimum cell design had a pressure sensor whose mechanical compliance was equal to the bulk compliance of the enclosed water. High sensitivity was achieved despite the use of lower power light sources, potentially extending the applicability of photothermal/photoacoustic methods to low-cost systems.

The photoacoustic cell incorporated a thin-walled piezoelectric cylinder. Using a 678nm laser diode emitting 1.4±0.2 mW rms, a minimum detectable absorption coefficient of 3x10-3cm-1 was demonstrated.

Better performance was achieved using the photothermal detector. Photothermal expansion in aqueous samples caused the deflection of a water meniscus held across a 200 μm-radius pinhole. Meniscus displacement was monitored using fibre optic interferometry. The magnitude and form of photothermal meniscus deflection signals were shown to be in good agreement with theory. Absorption by 1.5ppb anthracene in water was detected using a mercury discharge lamp (254nm) as an excitation source, and light emitting diodes were used to excite photothermal signals in a liquid (water) for the first time. The technique was limited by light source emission intensities and environmental acoustic noise, for example giving a minimum detectable absorption coefficient of 2x10-4cm-1 using an LED emitting 0.38±0.06 mW rms.

The minimum detectable absorption coefficients of the photothermal detector and of a conventional transmission spectrometer were found to be similar, but the photothermal technique was far less sensitive to the effects of light scattering. The ability to detect trace contaminants in water, in the presence of turbidity, is of great importance to the water industry.

This record has no associated files available for download.

More information

Published date: 1998

Identifiers

Local EPrints ID: 463286
URI: http://eprints.soton.ac.uk/id/eprint/463286
PURE UUID: 7a926288-010b-4f64-9527-150ef18c8ff6

Catalogue record

Date deposited: 04 Jul 2022 20:48
Last modified: 04 Jul 2022 20:48

Export record

Contributors

Author: Jane Hodgkinson

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×