Leadless pacemakers: where is the device?
Leadless pacemakers: where is the device?
Type of funding sources: None.
Introduction: Leadless pacemakers (LPs) were designed to avoid lead-related complications associated with transvenous pacing. To minimise the risk of complications, there is preference towards implanting LPs into the septal aspect of the right ventricle rather than the apex or free wall. The Transcatheter Pacing Study (TPS) and the international post-approval registry demonstrated the safety and reliability of the LP systems in real-world settings. The registry demonstrated that more than half of the LPs were implanted into the septum and most required <2 attempts at deployment. We report a radiological method of defining LP position.
Methods: We reviewed the first 100 LPs implanted at our centre. Two independent observers who didn’t implant LPs reviewed the patients’ post-implant fluoroscopy images and/or post-implant CXRs when available. The reviewers assessed the devices’ positions in postero-anterior (PA) and/or right anterior oblique (RAO) views based on conventional fluoroscopic criteria for lead position. We used the proposed criteria interchangeably on fluoroscopic images and post implant CXRs (Figure). Differences in classification of device position were resolved by consensus.
Results: Three experienced operators implanted 100 LPs at our centre. Patients (61% male) 56.6 ± 22.2 years had normal hearts (74%), ischaemic cardiomyopathies (12%), congenital heart diseases (6%), valvular pathologies (5%) and dilated cardiomyopathies (3%). Indications for pacing were symptomatic sinus node dysfunction (36%), followed by high grade atrio-ventricular block (33%), bradyarrhythmia associated with atrial tachyarrhythmias (27%) and other indications for pacing (4%). We had a 100% successful implant rate, 88% required ≤2 attempts and 70% required one attempt. There were no major complications.
We were able to classify the site of the LPs implants in a total of 90 patients who had fluoroscopic projections or chest x-rays that would allow us to classify the implant sites. A total of 32 implants were in the apex (35.6%). 28 were in mid-septum (31.1 %), 15 in the apical septum (16.7%), 14 on the septal aspect of the right ventricular inflow (15.5%) and 1 implant (1.1%) in the septum of the RV outflow tract.
Conclusion: Our proposed method of defining LP position demonstrated that the rate of implants into the true apex at our centre was highly comparable to that of the international registry. It also showed that we had lower rates of implants into the mid-septum in favour of apical septum. There were no pericardial effusions or cardiac perforations resulting from our implant procedures regardless of the site of the implant. We utilised widely used fluoroscopic and chest x-ray criteria for categorisation of the LPs implantation sites. However, a recognised limitation to our analysis is that our findings were not validated using other imaging modalities such as echocardiogram or cardiac computerised tomography (CT).
1099-5129
Elrefai, M
28916fea-4687-4d4b-99aa-961e73b710ab
Menexi, C
17fc132a-2c02-47ce-96de-436e5fe00675
Roberts, P
e544d250-471c-4b39-8f53-61ed24b54f2a
Elrefai, M
28916fea-4687-4d4b-99aa-961e73b710ab
Menexi, C
17fc132a-2c02-47ce-96de-436e5fe00675
Roberts, P
e544d250-471c-4b39-8f53-61ed24b54f2a

Elrefai, M, Menexi, C and Roberts, P (2021) Leadless pacemakers: where is the device? EP Europace, 23 (Supplement_3). (doi:10.1093/europace/euab116.375).

Record type: Article

Abstract

Type of funding sources: None.
Introduction: Leadless pacemakers (LPs) were designed to avoid lead-related complications associated with transvenous pacing. To minimise the risk of complications, there is preference towards implanting LPs into the septal aspect of the right ventricle rather than the apex or free wall. The Transcatheter Pacing Study (TPS) and the international post-approval registry demonstrated the safety and reliability of the LP systems in real-world settings. The registry demonstrated that more than half of the LPs were implanted into the septum and most required <2 attempts at deployment. We report a radiological method of defining LP position.
Methods: We reviewed the first 100 LPs implanted at our centre. Two independent observers who didn’t implant LPs reviewed the patients’ post-implant fluoroscopy images and/or post-implant CXRs when available. The reviewers assessed the devices’ positions in postero-anterior (PA) and/or right anterior oblique (RAO) views based on conventional fluoroscopic criteria for lead position. We used the proposed criteria interchangeably on fluoroscopic images and post implant CXRs (Figure). Differences in classification of device position were resolved by consensus.
Results: Three experienced operators implanted 100 LPs at our centre. Patients (61% male) 56.6 ± 22.2 years had normal hearts (74%), ischaemic cardiomyopathies (12%), congenital heart diseases (6%), valvular pathologies (5%) and dilated cardiomyopathies (3%). Indications for pacing were symptomatic sinus node dysfunction (36%), followed by high grade atrio-ventricular block (33%), bradyarrhythmia associated with atrial tachyarrhythmias (27%) and other indications for pacing (4%). We had a 100% successful implant rate, 88% required ≤2 attempts and 70% required one attempt. There were no major complications.
We were able to classify the site of the LPs implants in a total of 90 patients who had fluoroscopic projections or chest x-rays that would allow us to classify the implant sites. A total of 32 implants were in the apex (35.6%). 28 were in mid-septum (31.1 %), 15 in the apical septum (16.7%), 14 on the septal aspect of the right ventricular inflow (15.5%) and 1 implant (1.1%) in the septum of the RV outflow tract.
Conclusion: Our proposed method of defining LP position demonstrated that the rate of implants into the true apex at our centre was highly comparable to that of the international registry. It also showed that we had lower rates of implants into the mid-septum in favour of apical septum. There were no pericardial effusions or cardiac perforations resulting from our implant procedures regardless of the site of the implant. We utilised widely used fluoroscopic and chest x-ray criteria for categorisation of the LPs implantation sites. However, a recognised limitation to our analysis is that our findings were not validated using other imaging modalities such as echocardiogram or cardiac computerised tomography (CT).

This record has no associated files available for download.

More information

Published date: 24 May 2021

Identifiers

Local EPrints ID: 475818
URI: http://eprints.soton.ac.uk/id/eprint/475818
ISSN: 1099-5129
PURE UUID: fb244cc2-ea7b-47e6-a118-c9dc6e35094f

Catalogue record

Date deposited: 28 Mar 2023 18:36
Last modified: 17 Mar 2024 00:13

Export record

Altmetrics

Contributors

Author: M Elrefai
Author: C Menexi
Author: P Roberts

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×