Glenohumeral kinematics following total shoulder arthroplasty: a finite element investigation
Glenohumeral kinematics following total shoulder arthroplasty: a finite element investigation
The osseous geometry of the glenohumeral joint is naturally nonconforming and minimally constrained, and the joint’s stability is maintained by action of the rotator cuff muscles.
Damage to these muscles is often associated with joint degeneration, and a variety of glenoid prostheses have been developed to impart varying degrees of stability postoperatively. The issues of conformity and constraint within the artificial shoulder have been addressed through in vivo and in vitro studies, although few computational models have been presented. The current investigation presents the results of three-dimensional finite element analyses of the total shoulder joint and the effects of design parameters upon glenohumeral interaction. Conformity was shown not to influence
the loads required to destabilize the joint, although it was the principal factor determining the magnitude of humeral head translation. Constraint was found to correlate linearly with the
forces required to dislocate the humeral head, with higher constraint leading to slightly greater humeral migration at the point of joint instability. The model predicts that patients with a
dysfunctional supraspinatus would experience frequent eccentric loading of the glenoid, especially in the superior direction, which would likely lead to increased fixation stresses, and hence, a greater chance of loosening. For candidates with an intact rotator cuff, the models developed in this study
predict that angular constraints of at least 148 and 6.58 in the superoinferior and anteroposterior axes are required to provide stable unloaded abduction of the humerus, with larger constraints of 188 and 108 necessitated by a dysfunctional supraspinatus. The tools developed during this study can be
used to determine the capacity for different implant designs to provide resistance to excessive glenohumeral translations and reduce the potential for instability of the joint, allowing surgeons to optimize postoperative functional gains on a patient by patient basis.
shoulder, kinematics, tsa, finite element, arthroplasty
108-115
Hopkins, Andrew R.
5a263770-2eeb-4c09-a6e0-1e7b54435d8f
Hansen, Ulrich N.
ae9c69ab-87a8-4435-b113-ee319579adef
Amis, Andrew A.
31e15d47-8b4e-4279-a73b-f70038f90d48
Taylor, Mark
e368bda3-6ca5-4178-80e9-41a689badeeb
Emery, Roger J.
f443dd99-20e8-465e-b03c-8923a514ef1a
January 2007
Hopkins, Andrew R.
5a263770-2eeb-4c09-a6e0-1e7b54435d8f
Hansen, Ulrich N.
ae9c69ab-87a8-4435-b113-ee319579adef
Amis, Andrew A.
31e15d47-8b4e-4279-a73b-f70038f90d48
Taylor, Mark
e368bda3-6ca5-4178-80e9-41a689badeeb
Emery, Roger J.
f443dd99-20e8-465e-b03c-8923a514ef1a
Hopkins, Andrew R., Hansen, Ulrich N., Amis, Andrew A., Taylor, Mark and Emery, Roger J.
(2007)
Glenohumeral kinematics following total shoulder arthroplasty: a finite element investigation.
Journal of Orthopaedic Research, 25 (1), .
(doi:10.1002/jor.20290).
Abstract
The osseous geometry of the glenohumeral joint is naturally nonconforming and minimally constrained, and the joint’s stability is maintained by action of the rotator cuff muscles.
Damage to these muscles is often associated with joint degeneration, and a variety of glenoid prostheses have been developed to impart varying degrees of stability postoperatively. The issues of conformity and constraint within the artificial shoulder have been addressed through in vivo and in vitro studies, although few computational models have been presented. The current investigation presents the results of three-dimensional finite element analyses of the total shoulder joint and the effects of design parameters upon glenohumeral interaction. Conformity was shown not to influence
the loads required to destabilize the joint, although it was the principal factor determining the magnitude of humeral head translation. Constraint was found to correlate linearly with the
forces required to dislocate the humeral head, with higher constraint leading to slightly greater humeral migration at the point of joint instability. The model predicts that patients with a
dysfunctional supraspinatus would experience frequent eccentric loading of the glenoid, especially in the superior direction, which would likely lead to increased fixation stresses, and hence, a greater chance of loosening. For candidates with an intact rotator cuff, the models developed in this study
predict that angular constraints of at least 148 and 6.58 in the superoinferior and anteroposterior axes are required to provide stable unloaded abduction of the humerus, with larger constraints of 188 and 108 necessitated by a dysfunctional supraspinatus. The tools developed during this study can be
used to determine the capacity for different implant designs to provide resistance to excessive glenohumeral translations and reduce the potential for instability of the joint, allowing surgeons to optimize postoperative functional gains on a patient by patient basis.
This record has no associated files available for download.
More information
Published date: January 2007
Keywords:
shoulder, kinematics, tsa, finite element, arthroplasty
Identifiers
Local EPrints ID: 47586
URI: http://eprints.soton.ac.uk/id/eprint/47586
ISSN: 0736-0266
PURE UUID: dc611447-ed9a-4b74-9f5e-395668d0e611
Catalogue record
Date deposited: 03 Aug 2007
Last modified: 15 Mar 2024 09:34
Export record
Altmetrics
Contributors
Author:
Andrew R. Hopkins
Author:
Ulrich N. Hansen
Author:
Andrew A. Amis
Author:
Mark Taylor
Author:
Roger J. Emery
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics