Evaluating dynamic biomechanics at a transtibial socket interface during gait phases using finite element analysis

Evaluating dynamic biomechanics at a transtibial socket interface during gait phases using finite element analysis
Evaluating dynamic biomechanics at a transtibial socket interface during gait phases using finite element analysis
Introduction: the interface between a lower extremity amputee residuum and prosthetic socket is subjected to extensive loading during ambulation. Prolonged use of a prosthesis can result in discomfort and possible tissue injury, which is more likely with ill-fitted sockets. Study of the interface mechanical coupling is important to improve understanding at this important interface. Finite element analysis (FEA) has been extensively used to evaluate interface stresses and help assess load distribution and potential risk of tissue injuries [1]. However, few FEA studies report both pressure and shear at load-bearing locations, e.g., patella tendon (PT), fibula head (FH), popliteal depression (PD), especially at different phases of the gait cycle; nor utilise experiment outputs to assist validation. This work reports FEA pressure and shear of a transtibial (TT) interface during key gait phases such as weight acceptance (WA), midstance (MS) and toe-off (TO). Outputs were analysed alongside experimental sensor outputs obtained using a lab-based TT stump/socket simulator [2].

Methods: pressure and shear at the PT, FH and PD were experimentally measured using a stump/socket interface simulator. A corresponding FE model was created to align with the simulator’s set-up mimicking key gait phases, e.g., WA. Stump tissue was assigned a compressive modulus of 146kPa, aligning with the silicone material used in the simulator [2]. FEA pressure (P), circumferential shear (SC) and longitudinal shear (SL) at these load-bearing areas were obtained and validated by experimental sensor outputs.

Results & discussion: FEA results show peak pressures of 152kPa, 120kPa and 31kPa at PT, FH, and PD, respectively, during WA. Relatively higher P at PT is attributable to the bony prominence in the area as compared with FH and PD. Magnitude of SC at the three sites exhibited small differences (approximately 3kPa at PT, 1kPa at PD, and 4kPa at FH) which may suggest the stump’s negligible circumferential movement in the socket. SL towards the distal direction at FH (14kPa) and PT (36kPa) were observed, with no notable SL at PD. This aligns with the fact that the stump moves downwards in the socket during WA, as load is transferred from the contralateral to prosthetic side. The FEA outputs largely align well with interface pressure and shear sensor measurements under WA conditions when using the simulator. The validated model was further used to study pressure and shear during other gait phases including MS and TO. Detailed results and discussion on experimental data and FEA will be presented.

Conclusion: a computational FE model was developed to estimate pressure and shear distribution during typical phases in gait cycles. The model was evaluated using a corresponding simulator and interface sensor measurements.

References:[1] A. S. Dickinson, J. W. Steer, and P. R. Worsley, "Finite element analysis of the amputated lower limb: A systematic review and recommendations," Medical Engineering & Physics, vol. 43, pp. 1-18, 2017/05/01/ 2017, doi: https://doi.org/10.1016/j.medengphy.2017.02.008.
[2] M. P. McGrath et al., "Development of a residuum/socket interface simulator for lower limb prosthetics," Proc Inst Mech Eng H, vol. 231, no. 3, pp. 235-242, Mar 2017, doi: 10.1177/0954411917690764.



Devin, Kirstie
a8f23fa0-db53-44a4-abd8-03a72800f88d
Tang, Jinghua
b4b9a22c-fd6d-427a-9ab1-51184c1d2a2c
Moser, David
09874cab-348f-47f9-b018-1c2875d16998
Jiang, Liudi
374f2414-51f0-418f-a316-e7db0d6dc4d1
Devin, Kirstie
a8f23fa0-db53-44a4-abd8-03a72800f88d
Tang, Jinghua
b4b9a22c-fd6d-427a-9ab1-51184c1d2a2c
Moser, David
09874cab-348f-47f9-b018-1c2875d16998
Jiang, Liudi
374f2414-51f0-418f-a316-e7db0d6dc4d1

Devin, Kirstie, Tang, Jinghua, Moser, David and Jiang, Liudi (2022) Evaluating dynamic biomechanics at a transtibial socket interface during gait phases using finite element analysis. IPSO UL Annual Scientific Meeting & Exhibition, Village Hotel, London, United Kingdom. 06 - 07 Oct 2022. 1 pp .

Record type: Conference or Workshop Item (Other)

Abstract

Introduction: the interface between a lower extremity amputee residuum and prosthetic socket is subjected to extensive loading during ambulation. Prolonged use of a prosthesis can result in discomfort and possible tissue injury, which is more likely with ill-fitted sockets. Study of the interface mechanical coupling is important to improve understanding at this important interface. Finite element analysis (FEA) has been extensively used to evaluate interface stresses and help assess load distribution and potential risk of tissue injuries [1]. However, few FEA studies report both pressure and shear at load-bearing locations, e.g., patella tendon (PT), fibula head (FH), popliteal depression (PD), especially at different phases of the gait cycle; nor utilise experiment outputs to assist validation. This work reports FEA pressure and shear of a transtibial (TT) interface during key gait phases such as weight acceptance (WA), midstance (MS) and toe-off (TO). Outputs were analysed alongside experimental sensor outputs obtained using a lab-based TT stump/socket simulator [2].

Methods: pressure and shear at the PT, FH and PD were experimentally measured using a stump/socket interface simulator. A corresponding FE model was created to align with the simulator’s set-up mimicking key gait phases, e.g., WA. Stump tissue was assigned a compressive modulus of 146kPa, aligning with the silicone material used in the simulator [2]. FEA pressure (P), circumferential shear (SC) and longitudinal shear (SL) at these load-bearing areas were obtained and validated by experimental sensor outputs.

Results & discussion: FEA results show peak pressures of 152kPa, 120kPa and 31kPa at PT, FH, and PD, respectively, during WA. Relatively higher P at PT is attributable to the bony prominence in the area as compared with FH and PD. Magnitude of SC at the three sites exhibited small differences (approximately 3kPa at PT, 1kPa at PD, and 4kPa at FH) which may suggest the stump’s negligible circumferential movement in the socket. SL towards the distal direction at FH (14kPa) and PT (36kPa) were observed, with no notable SL at PD. This aligns with the fact that the stump moves downwards in the socket during WA, as load is transferred from the contralateral to prosthetic side. The FEA outputs largely align well with interface pressure and shear sensor measurements under WA conditions when using the simulator. The validated model was further used to study pressure and shear during other gait phases including MS and TO. Detailed results and discussion on experimental data and FEA will be presented.

Conclusion: a computational FE model was developed to estimate pressure and shear distribution during typical phases in gait cycles. The model was evaluated using a corresponding simulator and interface sensor measurements.

References:[1] A. S. Dickinson, J. W. Steer, and P. R. Worsley, "Finite element analysis of the amputated lower limb: A systematic review and recommendations," Medical Engineering & Physics, vol. 43, pp. 1-18, 2017/05/01/ 2017, doi: https://doi.org/10.1016/j.medengphy.2017.02.008.
[2] M. P. McGrath et al., "Development of a residuum/socket interface simulator for lower limb prosthetics," Proc Inst Mech Eng H, vol. 231, no. 3, pp. 235-242, Mar 2017, doi: 10.1177/0954411917690764.



Text
ISPOUKMS2022_Abstract_Final - Accepted Manuscript
Download (216kB)

More information

Published date: October 2022
Venue - Dates: IPSO UL Annual Scientific Meeting & Exhibition, Village Hotel, London, United Kingdom, 2022-10-06 - 2022-10-07

Identifiers

Local EPrints ID: 476610
URI: http://eprints.soton.ac.uk/id/eprint/476610
PURE UUID: a9561e7a-b94a-4052-ba10-46899688695a
ORCID for Kirstie Devin: ORCID iD orcid.org/0000-0001-6794-2375
ORCID for Jinghua Tang: ORCID iD orcid.org/0000-0003-3359-5891
ORCID for Liudi Jiang: ORCID iD orcid.org/0000-0002-3400-825X

Catalogue record

Date deposited: 10 May 2023 16:31
Last modified: 12 Nov 2024 03:16

Export record

Contributors

Author: Kirstie Devin ORCID iD
Author: Jinghua Tang ORCID iD
Author: David Moser
Author: Liudi Jiang ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×