The University of Southampton
University of Southampton Institutional Repository

Improving the sensitivity of cochlear implant integrity testing: recording electrode voltages with surface electrodes to diagnose partial short circuits in Advanced Bionics Ultra V1 devices

Improving the sensitivity of cochlear implant integrity testing: recording electrode voltages with surface electrodes to diagnose partial short circuits in Advanced Bionics Ultra V1 devices
Improving the sensitivity of cochlear implant integrity testing: recording electrode voltages with surface electrodes to diagnose partial short circuits in Advanced Bionics Ultra V1 devices
Background: advanced Bionics (AB) Ultra V1 devices are vulnerable to partial short circuits, resulting from moisture ingress. The manufacturer provides clinicians with a diagnostic test to identify this fault, “Electric Field Imaging” (EFI), using trans-impedance measurements. AB estimate the sensitivity of the algorithm incorporated in the test to be between 70 and 90%. We Recorded Electrode Voltages (REVs) using surface electrodes and compared these with the EFI test.

Method: 65 AB Ultra V1 devices supported by the University of Southampton Auditory Implant Service were assessed by REVs and EFI during routine clinic appointments. 32 devices had been tested at least twice, 16 in children and 16 in adults by March 2023. REVs was performed with surface electrodes placed on the earlobes and forehead, whilst the implant was simulated in live mode, taking approximately five minutes per ear.

Results: 10 of the 32 devices which had been tested at least twice presented as normal at the earlier test, based on the EFI test and two criteria previously suggested in the published literature for REVs. Relative Electrode Voltages (EVs) across the electrode array were calculated for measurements made with the
recording electrodes on the ipsilateral and contralateral earlobes. These were fitted to a polynomial (cubic) function, showing excellent agreement, mean R-squared=0.97 (range 0.89-0.997). For individual electrodes, the mean absolute difference between the predicted and measured values was 1.1%, st dev 1.3%. All REVs differed by Devices which presented as abnormal on the EFI test gave a poorer fit to a cubic function, mean Rsquared=0.79 (range 0.40-0.96), based on the later tests (N=33). The mean absolute difference between the predicted and measured EVs was 8.2%, st dev 9.3%. All devices had at least one electrode with an EV ≥9% different from the predicted value. 32 devices were returned as normal on the later EFI tests. Of these, 15 devices had REVs with a REV ≥9% different from the predicted value.

Conclusions: REVs was found to be highly sensitive to partial short circuits caused by moisture ingress in AB Ultra V1 devices. EVs were well fitted to a cubic function in normal devices. Results suggest that 77% of devices tested were affected by the fault at the later test.
cochlear implant
Grasmeder, Mary
206e6b44-d1cd-43f5-99ac-588ab02d44ef
Rogers, Katrine
54594b3d-b7b5-4f70-9269-853215db82eb
Aydin, Ziya
a59046ba-fde4-495b-ac92-a46a5391db6c
Hough, Kate
81d8630c-6e02-4bea-858a-377717476f6e
Verschuur, Carl
5e15ee1c-3a44-4dbe-ad43-ec3b50111e41
Newman, Tracey
322290cb-2e9c-445d-a047-00b1bea39a25
Grasmeder, Mary
206e6b44-d1cd-43f5-99ac-588ab02d44ef
Rogers, Katrine
54594b3d-b7b5-4f70-9269-853215db82eb
Aydin, Ziya
a59046ba-fde4-495b-ac92-a46a5391db6c
Hough, Kate
81d8630c-6e02-4bea-858a-377717476f6e
Verschuur, Carl
5e15ee1c-3a44-4dbe-ad43-ec3b50111e41
Newman, Tracey
322290cb-2e9c-445d-a047-00b1bea39a25

Grasmeder, Mary, Rogers, Katrine, Aydin, Ziya, Hough, Kate, Verschuur, Carl and Newman, Tracey (2023) Improving the sensitivity of cochlear implant integrity testing: recording electrode voltages with surface electrodes to diagnose partial short circuits in Advanced Bionics Ultra V1 devices. 11th International Symposium on Objective Measures in auditory implants, Flanders Meeting & Convention Centre, Antwerp, Belgium. 25 - 28 Sep 2023.

Record type: Conference or Workshop Item (Paper)

Abstract

Background: advanced Bionics (AB) Ultra V1 devices are vulnerable to partial short circuits, resulting from moisture ingress. The manufacturer provides clinicians with a diagnostic test to identify this fault, “Electric Field Imaging” (EFI), using trans-impedance measurements. AB estimate the sensitivity of the algorithm incorporated in the test to be between 70 and 90%. We Recorded Electrode Voltages (REVs) using surface electrodes and compared these with the EFI test.

Method: 65 AB Ultra V1 devices supported by the University of Southampton Auditory Implant Service were assessed by REVs and EFI during routine clinic appointments. 32 devices had been tested at least twice, 16 in children and 16 in adults by March 2023. REVs was performed with surface electrodes placed on the earlobes and forehead, whilst the implant was simulated in live mode, taking approximately five minutes per ear.

Results: 10 of the 32 devices which had been tested at least twice presented as normal at the earlier test, based on the EFI test and two criteria previously suggested in the published literature for REVs. Relative Electrode Voltages (EVs) across the electrode array were calculated for measurements made with the
recording electrodes on the ipsilateral and contralateral earlobes. These were fitted to a polynomial (cubic) function, showing excellent agreement, mean R-squared=0.97 (range 0.89-0.997). For individual electrodes, the mean absolute difference between the predicted and measured values was 1.1%, st dev 1.3%. All REVs differed by Devices which presented as abnormal on the EFI test gave a poorer fit to a cubic function, mean Rsquared=0.79 (range 0.40-0.96), based on the later tests (N=33). The mean absolute difference between the predicted and measured EVs was 8.2%, st dev 9.3%. All devices had at least one electrode with an EV ≥9% different from the predicted value. 32 devices were returned as normal on the later EFI tests. Of these, 15 devices had REVs with a REV ≥9% different from the predicted value.

Conclusions: REVs was found to be highly sensitive to partial short circuits caused by moisture ingress in AB Ultra V1 devices. EVs were well fitted to a cubic function in normal devices. Results suggest that 77% of devices tested were affected by the fault at the later test.

This record has no associated files available for download.

More information

Published date: 26 September 2023
Venue - Dates: 11th International Symposium on Objective Measures in auditory implants, Flanders Meeting & Convention Centre, Antwerp, Belgium, 2023-09-25 - 2023-09-28
Keywords: cochlear implant

Identifiers

Local EPrints ID: 486029
URI: http://eprints.soton.ac.uk/id/eprint/486029
PURE UUID: 41e882b0-f760-40f6-a386-307bd26b9455
ORCID for Kate Hough: ORCID iD orcid.org/0000-0002-5160-2517
ORCID for Tracey Newman: ORCID iD orcid.org/0000-0002-3727-9258

Catalogue record

Date deposited: 05 Jan 2024 18:06
Last modified: 30 Nov 2024 03:08

Export record

Contributors

Author: Mary Grasmeder
Author: Katrine Rogers
Author: Ziya Aydin
Author: Kate Hough ORCID iD
Author: Carl Verschuur
Author: Tracey Newman ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×