Evaluation of the Aanderaa Oxygen Optode in continuous use in the NOC Portsmouth Bilbao FerryBox system 2005, 2006, with an assessment of the likely errors in the estimation of oxygen concentration anomalies
Evaluation of the Aanderaa Oxygen Optode in continuous use in the NOC Portsmouth Bilbao FerryBox system 2005, 2006, with an assessment of the likely errors in the estimation of oxygen concentration anomalies
An Aanderaa Oxygen Optode has been used as part of the “FerryBox” system on the P&O Ferries vessel the MV Pride of Bilbao since the beginning of 2005. This report covers data collected through 2005 and 2006. The accuracy, precision and stability of the Optode have been compared with (1) measurements of oxygen in water samples collected and processed on board the ship by Winkler titration and (2) estimates of the oxygen anomaly. The comparison of the Winkler data with the output from the Optode suggests that the accuracy of the Optode in both years was close to 98% of the true value when compared to the Winkler measurements. The data suggest that the precision based on the comparison of the two data sets was equivalent to a standard deviation of 5 ?M or 2% of the average value in 2005 and 2.9 or 1% of the average value in 2006. The resolution of the recorded a data is better than 1 ?M.
Two sources of error in measurement of temperature which are significant for the calculation of the oxygen anomaly have been identified. Temperatures measured using sensors in board the ship are higher than “true” temperatures. The error changes through the year as the difference in temperature between inside the ship and the external water temperature changes. This error may be as large as 0.5 °C. The anomaly has to be calculated using temperature measurement in the water flowing past the Optode. This is because of delays and mixing in the water being pumped into the ship have smoothing effect on the data, and can enhance the difference between the internal and external temperature to being as high as 1.0 °C. Consequently to avoid this large error due to the timing of the measurements the oxygen anomaly should be calculated from the temperature measured in the pumped water stream adjusted by an offset to the “true” outside temperature.
Bay of Biscay, dissolved oxygen, English Channel, FerryBox, gas exchange, monitoring, Optode, productivity, seawater, Winkler titration
National Oceanography Centre
Hydes, D.J.
ac7371d4-c2b9-4926-bb77-ce58480ecff7
Hartman, M.C.
cdd63b60-f89a-4ef8-842d-7803f8213c18
Hartman, S.E.
2f74a439-395a-4ee7-89a2-eff4cc8d9481
Bargeron, C.P.
14a61c9c-b684-450c-b127-340e6cbe734f
October 2007
Hydes, D.J.
ac7371d4-c2b9-4926-bb77-ce58480ecff7
Hartman, M.C.
cdd63b60-f89a-4ef8-842d-7803f8213c18
Hartman, S.E.
2f74a439-395a-4ee7-89a2-eff4cc8d9481
Bargeron, C.P.
14a61c9c-b684-450c-b127-340e6cbe734f
Hydes, D.J., Hartman, M.C., Hartman, S.E. and Bargeron, C.P.
(2007)
Evaluation of the Aanderaa Oxygen Optode in continuous use in the NOC Portsmouth Bilbao FerryBox system 2005, 2006, with an assessment of the likely errors in the estimation of oxygen concentration anomalies
(National Oceanography Centre Southampton Internal Document, 7)
Southampton, UK.
National Oceanography Centre
73pp.
Record type:
Monograph
(Project Report)
Abstract
An Aanderaa Oxygen Optode has been used as part of the “FerryBox” system on the P&O Ferries vessel the MV Pride of Bilbao since the beginning of 2005. This report covers data collected through 2005 and 2006. The accuracy, precision and stability of the Optode have been compared with (1) measurements of oxygen in water samples collected and processed on board the ship by Winkler titration and (2) estimates of the oxygen anomaly. The comparison of the Winkler data with the output from the Optode suggests that the accuracy of the Optode in both years was close to 98% of the true value when compared to the Winkler measurements. The data suggest that the precision based on the comparison of the two data sets was equivalent to a standard deviation of 5 ?M or 2% of the average value in 2005 and 2.9 or 1% of the average value in 2006. The resolution of the recorded a data is better than 1 ?M.
Two sources of error in measurement of temperature which are significant for the calculation of the oxygen anomaly have been identified. Temperatures measured using sensors in board the ship are higher than “true” temperatures. The error changes through the year as the difference in temperature between inside the ship and the external water temperature changes. This error may be as large as 0.5 °C. The anomaly has to be calculated using temperature measurement in the water flowing past the Optode. This is because of delays and mixing in the water being pumped into the ship have smoothing effect on the data, and can enhance the difference between the internal and external temperature to being as high as 1.0 °C. Consequently to avoid this large error due to the timing of the measurements the oxygen anomaly should be calculated from the temperature measured in the pumped water stream adjusted by an offset to the “true” outside temperature.
Text
NOCS_ID_7(3).pdf
- Other
More information
Published date: October 2007
Additional Information:
Internal Document deposted at request of D.J. Hydes
Keywords:
Bay of Biscay, dissolved oxygen, English Channel, FerryBox, gas exchange, monitoring, Optode, productivity, seawater, Winkler titration
Identifiers
Local EPrints ID: 48673
URI: http://eprints.soton.ac.uk/id/eprint/48673
PURE UUID: ba01ea06-8791-40c8-b378-6ad38fb69b7e
Catalogue record
Date deposited: 05 Oct 2007
Last modified: 09 Apr 2024 16:34
Export record
Contributors
Author:
D.J. Hydes
Author:
M.C. Hartman
Author:
S.E. Hartman
Author:
C.P. Bargeron
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.