Objective: Adults typically receive only one cochlear implant (CI) due to cost constraints, with a contralateral hearing aid recommended when there is aidable hearing. Standard hearing aids differ from a CI in terms of processing strategy and function as a separate entity, requiring the user to integrate the disparate signals. Integrated bimodal technology has recently been introduced to address this challenge. The aim of the study was to investigate the performance of unilateral CI users with and without an integrated bimodal fitting and determine whether binaural streaming technology offers additional benefit.
Study sample: Twenty-six CI users using integrated bimodal technology. Design: Repeated measures where outcomes and user experience were assessed using a functional test battery more representative of real life listening (speech perception in noise tests, localisation test, tracking test) and the speech, spatial and qualities-of-hearing scale (SSQ).
Results: Bimodal outcomes were significantly better than for CI alone. Speech perception in noise improvements ranged from 1.4 dB to 3.5 dB depending on the location of speech and noise. The localisation and tracking tests, and the SSQ also showed significant improvements. Binaural streaming offered
additional improvement (1.2 dB to 6.1 dB on the different speech tests).
Conclusions: Integrated bimodal and binaural streaming technology improved the performance of unilateral CI users.
Cochlear implant; hearing aid; integrated bimodal technology; real-life tests, real-life tests, hearing aid, Cochlear implant, integrated bimodal technology
1499-2027
Alfakhri, Manal
dbba8db1-371c-4a2f-afe1-425046dc827c
Campbell, Nicole
fde07dcf-4874-4bab-ab3a-c3bc3c0015da
Lineton, Ben
1ace4e96-34da-4fc4-bc17-a1d82b2ba0e2
Verschuur, Carl
5e15ee1c-3a44-4dbe-ad43-ec3b50111e41
Alfakhri, Manal
dbba8db1-371c-4a2f-afe1-425046dc827c
Campbell, Nicole
fde07dcf-4874-4bab-ab3a-c3bc3c0015da
Lineton, Ben
1ace4e96-34da-4fc4-bc17-a1d82b2ba0e2
Verschuur, Carl
5e15ee1c-3a44-4dbe-ad43-ec3b50111e41

Alfakhri, Manal, Campbell, Nicole, Lineton, Ben and Verschuur, Carl (2024) Integrated bimodal fitting and binaural streaming technology outcomes for unilateral cochlear implant users. International Journal of Audiology. (doi:10.1080/14992027.2024.2341954).

Record type: Article

Abstract

Objective: Adults typically receive only one cochlear implant (CI) due to cost constraints, with a contralateral hearing aid recommended when there is aidable hearing. Standard hearing aids differ from a CI in terms of processing strategy and function as a separate entity, requiring the user to integrate the disparate signals. Integrated bimodal technology has recently been introduced to address this challenge. The aim of the study was to investigate the performance of unilateral CI users with and without an integrated bimodal fitting and determine whether binaural streaming technology offers additional benefit.
Study sample: Twenty-six CI users using integrated bimodal technology. Design: Repeated measures where outcomes and user experience were assessed using a functional test battery more representative of real life listening (speech perception in noise tests, localisation test, tracking test) and the speech, spatial and qualities-of-hearing scale (SSQ).
Results: Bimodal outcomes were significantly better than for CI alone. Speech perception in noise improvements ranged from 1.4 dB to 3.5 dB depending on the location of speech and noise. The localisation and tracking tests, and the SSQ also showed significant improvements. Binaural streaming offered
additional improvement (1.2 dB to 6.1 dB on the different speech tests).
Conclusions: Integrated bimodal and binaural streaming technology improved the performance of unilateral CI users.

Text
Integrated bimodal fitting and binaural streaming technology outcomes for unilateral cochlear implant users - Version of Record
Available under License Creative Commons Attribution.
Download (1MB)

More information

Accepted/In Press date: 5 April 2024
e-pub ahead of print date: 3 May 2024
Published date: 3 May 2024
Additional Information: Publisher Copyright: © 2024 The Author(s). Published by Informa UK Limited, trading as Taylor & Francis Group on behalf of British Society of Audiology, International Society of Audiology, and Nordic Audiological Society.
Keywords: Cochlear implant; hearing aid; integrated bimodal technology; real-life tests, real-life tests, hearing aid, Cochlear implant, integrated bimodal technology

Identifiers

Local EPrints ID: 489979
URI: http://eprints.soton.ac.uk/id/eprint/489979
ISSN: 1499-2027
PURE UUID: 7e87d190-e8de-4f89-88c0-875fa6d234f1
ORCID for Manal Alfakhri: ORCID iD orcid.org/0000-0003-1837-4863
ORCID for Nicole Campbell: ORCID iD orcid.org/0000-0001-6895-5434
ORCID for Ben Lineton: ORCID iD orcid.org/0000-0003-4784-7762

Catalogue record

Date deposited: 09 May 2024 16:35
Last modified: 05 Jun 2024 01:39

Export record

Altmetrics

Contributors

Author: Manal Alfakhri ORCID iD
Author: Nicole Campbell ORCID iD
Author: Ben Lineton ORCID iD
Author: Carl Verschuur

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×