Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump

Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump
Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump

Aims: Impeller design has a significant impact on the overall performance of a blood pump. In this study, the effect of the blade curvature was investigated by performing in silico and in vitro studies on a recently developed centrifugal blood pump. Methods: A computational fluid dynamics study was performed for the flow rates of 3-5 L/min at 2000 r/min. The computational fluid dynamics model was also applied on the US Food and Drug Administration (FDA) benchmark blood pump to validate our computational method. The relative hemolysis index was calculated with the Eulerian hemolysis estimation method for five impellers with the wrap angles ranging from 0° to 240°. Hydraulic experiments were conducted for the validation of computational fluid dynamics results. In addition, the curved-blade impeller (120°) and the straight-blade impeller (0°) were evaluated with in vitro hemolysis tests using human blood. Results: The wrap angle of 120° provided the best hydraulic and hemolytic performance. Pump achieved the physiologic operating pressures and flows with 85-115 mmHg at 2.5-5.9 L/min. Compared to the straight-blade impeller, the 120° model reduces the relative hemolysis index and the plasma-free hemoglobin near 72.8% and 56.7%, respectively. Comparison of in silico and in vitro results indicated the similar trend to the blade curvature. Conclusion: Introducing a blade curvature enhanced the hydrodynamic and hemolytic performance compared to the straight-blade configuration for the investigated centrifugal blood pump. The findings of this study provide new insights into centrifugal blood pump design by examining the influence of the blade curvature.

Blade curvature, Centrifugal blood pump, Computational fluid dynamics, Hemolysis estimation, Impeller design
0391-3988
730-737
Ozturk, Caglar
70bbd3bd-fc56-48e8-8b5e-00d5270c1526
Aka, I. Basar
5178df04-8b70-42a5-9ede-3f2568c683a8
Lazoglu, Ismail
dde74a90-ccd3-452a-a4cf-6dbc8841caf3
Ozturk, Caglar
70bbd3bd-fc56-48e8-8b5e-00d5270c1526
Aka, I. Basar
5178df04-8b70-42a5-9ede-3f2568c683a8
Lazoglu, Ismail
dde74a90-ccd3-452a-a4cf-6dbc8841caf3

Ozturk, Caglar, Aka, I. Basar and Lazoglu, Ismail (2018) Effect of blade curvature on the hemolytic and hydraulic characteristics of a centrifugal blood pump. International Journal of Artificial Organs, 41 (11), 730-737. (doi:10.1177/0391398818785558).

Record type: Article

Abstract

Aims: Impeller design has a significant impact on the overall performance of a blood pump. In this study, the effect of the blade curvature was investigated by performing in silico and in vitro studies on a recently developed centrifugal blood pump. Methods: A computational fluid dynamics study was performed for the flow rates of 3-5 L/min at 2000 r/min. The computational fluid dynamics model was also applied on the US Food and Drug Administration (FDA) benchmark blood pump to validate our computational method. The relative hemolysis index was calculated with the Eulerian hemolysis estimation method for five impellers with the wrap angles ranging from 0° to 240°. Hydraulic experiments were conducted for the validation of computational fluid dynamics results. In addition, the curved-blade impeller (120°) and the straight-blade impeller (0°) were evaluated with in vitro hemolysis tests using human blood. Results: The wrap angle of 120° provided the best hydraulic and hemolytic performance. Pump achieved the physiologic operating pressures and flows with 85-115 mmHg at 2.5-5.9 L/min. Compared to the straight-blade impeller, the 120° model reduces the relative hemolysis index and the plasma-free hemoglobin near 72.8% and 56.7%, respectively. Comparison of in silico and in vitro results indicated the similar trend to the blade curvature. Conclusion: Introducing a blade curvature enhanced the hydrodynamic and hemolytic performance compared to the straight-blade configuration for the investigated centrifugal blood pump. The findings of this study provide new insights into centrifugal blood pump design by examining the influence of the blade curvature.

This record has no associated files available for download.

More information

e-pub ahead of print date: 12 July 2018
Additional Information: Publisher Copyright: © The Author(s) 2018.
Keywords: Blade curvature, Centrifugal blood pump, Computational fluid dynamics, Hemolysis estimation, Impeller design

Identifiers

Local EPrints ID: 490731
URI: http://eprints.soton.ac.uk/id/eprint/490731
ISSN: 0391-3988
PURE UUID: 35ff67e7-3279-42a3-9434-9c4cd994e0dd
ORCID for Caglar Ozturk: ORCID iD orcid.org/0000-0002-3688-0148

Catalogue record

Date deposited: 04 Jun 2024 17:00
Last modified: 05 Jun 2024 02:09

Export record

Altmetrics

Contributors

Author: Caglar Ozturk ORCID iD
Author: I. Basar Aka
Author: Ismail Lazoglu

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×