Comparison of curvilinear and linear ultrasound imaging probes for measuring cross-sectional area and linear dimensions

Comparison of curvilinear and linear ultrasound imaging probes for measuring cross-sectional area and linear dimensions
Comparison of curvilinear and linear ultrasound imaging probes for measuring cross-sectional area and linear dimensions
The aim of the study was to determine whether different ultrasound probe/transducer configurations produce the same measurements. Two investigators undertook 10 scans of a general purpose semi-solid multi-tissue ultrasound phantom (phantom A) using two ultrasound scanners with a linear and curvilinear probe. From those 10 scans, two measurements of cross-sectional area (CSA), width and thickness were made. These measurements were then repeated with an open-top fluid-filled phantom, with 10% ethanol solution (phantom B). Intra- and inter-rater reliability were examined using Bland and Altman plots. Agreement between measurements made with the two probe types was also assessed using Bland and Altman plots. An independent samples t-test was used to compare statistical differences between probe type configuration. There was a significant difference (p < 0.05) and a tendency for increased measurements in CSA and width, and decreased measurements in thickness when using a curvilinear probe on phantom A, and these differences were not scanner specific. When imaging phantom B there were no significant differences in measurements between probe configurations; however there was a small bias for smaller CSA measurements with a curvilinear probe. In conclusion there are small differences in measurements obtained from different ultrasound probe configurations using a semi-solid phantom, but their clinical significance is unknown.
ultrasound imaging, validation study, accuracy, probe configuration, phantom
0309-1902
498-504
Warner, M.B.
f4dce73d-fb87-4f71-a3f0-078123aa040c
Cotton, Angela
baa16dcd-cd16-40d7-8ab8-af816bc4ee1a
Stokes, M.J.
71730503-70ce-4e67-b7ea-a3e54579717f
Warner, M.B.
f4dce73d-fb87-4f71-a3f0-078123aa040c
Cotton, Angela
baa16dcd-cd16-40d7-8ab8-af816bc4ee1a
Stokes, M.J.
71730503-70ce-4e67-b7ea-a3e54579717f

Warner, M.B., Cotton, Angela and Stokes, M.J. (2007) Comparison of curvilinear and linear ultrasound imaging probes for measuring cross-sectional area and linear dimensions. Journal of Medical Engineering & Technology, 32 (6), 498-504. (doi:10.1080/03091900701695533).

Record type: Article

Abstract

The aim of the study was to determine whether different ultrasound probe/transducer configurations produce the same measurements. Two investigators undertook 10 scans of a general purpose semi-solid multi-tissue ultrasound phantom (phantom A) using two ultrasound scanners with a linear and curvilinear probe. From those 10 scans, two measurements of cross-sectional area (CSA), width and thickness were made. These measurements were then repeated with an open-top fluid-filled phantom, with 10% ethanol solution (phantom B). Intra- and inter-rater reliability were examined using Bland and Altman plots. Agreement between measurements made with the two probe types was also assessed using Bland and Altman plots. An independent samples t-test was used to compare statistical differences between probe type configuration. There was a significant difference (p < 0.05) and a tendency for increased measurements in CSA and width, and decreased measurements in thickness when using a curvilinear probe on phantom A, and these differences were not scanner specific. When imaging phantom B there were no significant differences in measurements between probe configurations; however there was a small bias for smaller CSA measurements with a curvilinear probe. In conclusion there are small differences in measurements obtained from different ultrasound probe configurations using a semi-solid phantom, but their clinical significance is unknown.

This record has no associated files available for download.

More information

Published date: 3 December 2007
Keywords: ultrasound imaging, validation study, accuracy, probe configuration, phantom

Identifiers

Local EPrints ID: 55258
URI: http://eprints.soton.ac.uk/id/eprint/55258
ISSN: 0309-1902
PURE UUID: d0e671e6-d26f-4dca-a112-915adf671014
ORCID for M.B. Warner: ORCID iD orcid.org/0000-0002-1483-0561
ORCID for M.J. Stokes: ORCID iD orcid.org/0000-0002-4204-0890

Catalogue record

Date deposited: 15 Aug 2008
Last modified: 16 Mar 2024 03:40

Export record

Altmetrics

Contributors

Author: M.B. Warner ORCID iD
Author: Angela Cotton
Author: M.J. Stokes ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×