Respiratory rate is an important physiological parameter, and its continuous monitoring plays a vital role in health management. This study investigates the application of a planar capacitive sensor equipped with active shielding, the capaciflector, in monitoring respiratory activity. Previous studies have assumed that changes in sensor capacitance are mainly caused by skin deformation during breathing, but this theory has not been experimentally confirmed. In this study, an integrated wireless capaciflector-based respiratory sensor was developed based on the Texas Instruments' FDC2214 capacitance-to-digital converter (CDC). Experimental tests in 10 adult male volunteers showed that skin stretching led to an increase in sensor capacitance, confirming that skin stretching is the main driver of the respiratory signal. In addition, the study examined the effect of the operating frequency (0.23-7.41 MHz) on sensor sensitivity. The sensor sensitivity at an operating frequency of 230 kHz was found to be 1.2 times higher than that at other experimental frequencies, which has the potential to be used as an optimized operating frequency. The study also explored the effect of different electrode sizes in the capaciflector on its sensitivity. The results showed that the change in electrode size did not significantly affect the sensitivity of the sensor, but the design with electrode length of 16 mm and spacing of 4 mm as well as electrode length of 12 mm and spacing of 2 mm showed relatively high sensitivity in some subjects, which may become a potential optimized size design. This study provides an important reference for the design of a respiratory sensor based on a capaciflector.
Active shielding, capacitive sensor, respiration measurement
39867-39880
Zhao, Boyuan
34031ccf-1442-4b19-9e7f-874a2e2c4770
Tyson, Jack J.
8ac7797a-1914-48d4-be2c-a8c76384ec80
Spencer, Daniel C.
4affe9f6-353a-4507-8066-0180b8dc9eaf
White, Neil M.
c7be4c26-e419-4e5c-9420-09fc02e2ac9c
1 November 2025
Zhao, Boyuan
34031ccf-1442-4b19-9e7f-874a2e2c4770
Tyson, Jack J.
8ac7797a-1914-48d4-be2c-a8c76384ec80
Spencer, Daniel C.
4affe9f6-353a-4507-8066-0180b8dc9eaf
White, Neil M.
c7be4c26-e419-4e5c-9420-09fc02e2ac9c
Zhao, Boyuan, Tyson, Jack J., Spencer, Daniel C. and White, Neil M.
(2025)
Comprehensive study of capaciflector-based respiration sensor performance: measurement principle and influence of operating frequency and electrode size.
IEEE Sensors Journal, 25 (21), .
(doi:10.1109/JSEN.2025.3612725).
Abstract
Respiratory rate is an important physiological parameter, and its continuous monitoring plays a vital role in health management. This study investigates the application of a planar capacitive sensor equipped with active shielding, the capaciflector, in monitoring respiratory activity. Previous studies have assumed that changes in sensor capacitance are mainly caused by skin deformation during breathing, but this theory has not been experimentally confirmed. In this study, an integrated wireless capaciflector-based respiratory sensor was developed based on the Texas Instruments' FDC2214 capacitance-to-digital converter (CDC). Experimental tests in 10 adult male volunteers showed that skin stretching led to an increase in sensor capacitance, confirming that skin stretching is the main driver of the respiratory signal. In addition, the study examined the effect of the operating frequency (0.23-7.41 MHz) on sensor sensitivity. The sensor sensitivity at an operating frequency of 230 kHz was found to be 1.2 times higher than that at other experimental frequencies, which has the potential to be used as an optimized operating frequency. The study also explored the effect of different electrode sizes in the capaciflector on its sensitivity. The results showed that the change in electrode size did not significantly affect the sensitivity of the sensor, but the design with electrode length of 16 mm and spacing of 4 mm as well as electrode length of 12 mm and spacing of 2 mm showed relatively high sensitivity in some subjects, which may become a potential optimized size design. This study provides an important reference for the design of a respiratory sensor based on a capaciflector.
This record has no associated files available for download.
More information
e-pub ahead of print date: 26 September 2025
Published date: 1 November 2025
Keywords:
Active shielding, capacitive sensor, respiration measurement
Identifiers
Local EPrints ID: 513306
URI: https://eprints.soton.ac.uk/id/eprint/513306
ISSN: 1530-437X
PURE UUID: 4d1da1ec-7aa8-4a26-8d3b-bf6e51b898e4
Catalogue record
Date deposited: 11 Aug 2026 16:53
Last modified: 13 Sep 2026 15:50
Export record
Altmetrics
Contributors
Author:
Boyuan Zhao
Author:
Jack J. Tyson
Author:
Daniel C. Spencer
Author:
Neil M. White
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics