An experimental analysis of thick-film solid-state reference electrodes


Sophocleous, Marios, Glanc-Gostkiewicz, Monika, Atkinson, John Karl and Garcia-Breijo, Eduardo (2012) An experimental analysis of thick-film solid-state reference electrodes. At IEEE Sensors 2012, Taipei, TW, 28 - 31 Oct 2012. 4pp.

Download

[img]
Preview
PDF
Download (474Kb)
[img]
Preview
PDF
Download (887Kb)

Description/Abstract

Thick-Film planar solid-state Silver/Silver Chloride (Ag/AgCl) reference electrodes were developed and tested for ion susceptibility and long term drift in approximately 0.04 M potassium chloride (KCl) solution. Various types of electrodes were tested exhibiting stabilities down to 2 millivolts per decade change of chloride concentration. It is demonstrated that Thick-Film reference electrodes are suitable for use in underground soil measurements due to their ruggedness and robustness.