Performance of miniaturised thick-film solid state pH sensors
Performance of miniaturised thick-film solid state pH sensors
The performance of novel solid-state Thick-Film pH sensor for water quality sampling suitable for deployment in remote catchment areas is described. These miniaturised screen printed planar pH sensors are an alternative to glass pH electrodes, which have many disadvantages such as high-cost, large size, mechanical fragility and limited shape. One of the approaches for improvement of the ruggedness of the pH electrode is the implementation of metal oxides as ion selective electrodes used in combination with screen printed silver/silver chloride (Ag/AgCl) reference electrodes. Various materials of Thick-Film pH sensors have been evaluated, these give a sensitivity close to the theoretical Nernstian value of 59 mV/pH at 25ºC.
1299-1302
Glanc-Gostkiewicz, Monika
fde22742-587f-4688-9f3a-2596457c1669
Sophocleous, Marios
908541f4-d6e1-4fdf-9423-fddc2b12bcbf
Atkinson, John Karl
5e9729b2-0e1f-400d-a889-c74f6390ea58
Garcia Breijo, Eduardo
12887272-e31e-43f3-a6f6-1072abcdb4f5
Glanc-Gostkiewicz, Monika
fde22742-587f-4688-9f3a-2596457c1669
Sophocleous, Marios
908541f4-d6e1-4fdf-9423-fddc2b12bcbf
Atkinson, John Karl
5e9729b2-0e1f-400d-a889-c74f6390ea58
Garcia Breijo, Eduardo
12887272-e31e-43f3-a6f6-1072abcdb4f5
Glanc-Gostkiewicz, Monika, Sophocleous, Marios, Atkinson, John Karl and Garcia Breijo, Eduardo
(2012)
Performance of miniaturised thick-film solid state pH sensors.
Procedia Engineering, 47, .
(doi:10.1016/j.proeng.2012.09.393).
Abstract
The performance of novel solid-state Thick-Film pH sensor for water quality sampling suitable for deployment in remote catchment areas is described. These miniaturised screen printed planar pH sensors are an alternative to glass pH electrodes, which have many disadvantages such as high-cost, large size, mechanical fragility and limited shape. One of the approaches for improvement of the ruggedness of the pH electrode is the implementation of metal oxides as ion selective electrodes used in combination with screen printed silver/silver chloride (Ag/AgCl) reference electrodes. Various materials of Thick-Film pH sensors have been evaluated, these give a sensitivity close to the theoretical Nernstian value of 59 mV/pH at 25ºC.
Text
EUROSENSORS_2012__Monika_Glanc-Gostkiewicz.pdf
- Accepted Manuscript
Restricted to Repository staff only
Request a copy
Text
1-s2.0-S1877705812044566-main
- Version of Record
Slideshow
Eurosensors_2012_poster_final.pptx
- Other
More information
e-pub ahead of print date: 2 November 2012
Venue - Dates:
The 26th European Conference on Solid-state Transducers (Eurosensors XXVI), Poland, Republic of, Poland, 2012-09-09 - 2012-09-12
Organisations:
Mechatronics
Identifiers
Local EPrints ID: 345486
URI: http://eprints.soton.ac.uk/id/eprint/345486
PURE UUID: 0afee18b-06ec-4dfd-9a64-61894cde6188
Catalogue record
Date deposited: 23 Nov 2012 09:56
Last modified: 15 Mar 2024 02:33
Export record
Altmetrics
Contributors
Author:
Monika Glanc-Gostkiewicz
Author:
Marios Sophocleous
Author:
Eduardo Garcia Breijo
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics