Does a PEEK femoral TKA implant preserve intact femoral surface strains compared with CoCr? A preliminary laboratory study

Does a PEEK femoral TKA implant preserve intact femoral surface strains compared with CoCr? A preliminary laboratory study
Does a PEEK femoral TKA implant preserve intact femoral surface strains compared with CoCr? A preliminary laboratory study
BACKGROUND: Both the material and geometry of a total knee arthroplasty (TKA) component influence the induced periprosthetic bone strain field. Strain, a measure of the local relative deformation in a structure, corresponds to the mechanical stimulus that governs bone remodeling and is therefore a useful in vitro biomechanical measure for assessing the response of bone to new implant designs and materials. A polyetheretherketone (PEEK) femoral implant has the potential to promote bone strains closer to that of natural bone as a result of its low elastic modulus compared with cobalt-chromium (CoCr).

QUESTIONS/PURPOSES: In the present study, we used a Digital Image Correlation (DIC) technique to answer the following question: Does a PEEK TKA femoral component induce a more physiologically normal bone strain distribution than a CoCr component? To achieve this, a DIC test protocol was developed for periprosthetic bone strain assessment using an analog model; the protocol aimed to minimize errors in strain assessment through the selection of appropriate analysis parameters.

METHODS: Three synthetic bone femurs were used in this experiment. One was implanted with a CoCr femoral component and one with a PEEK femoral component. The third (unimplanted) femur was intact and used as the physiological reference (control) model. All models were subjected to standing loads on the corresponding polyethylene (ultrahigh-molecular-weight polyethylene) tibial component, and speckle image data were acquired for surface strain analysis using DIC in six repeat tests. The strain in 16 regions of interest on the lateral surface of each of the implanted bone models was plotted for comparison with the corresponding strains in the intact case. A Wilcoxon signed-rank test was used to test for difference at the 5% significance level.

RESULTS: Surface analog bone strain after CoCr implantation indicated strain shielding (R2 = 0.6178 with slope, ? = 0.4314) and was lower than the intact case (p = 0.014). The strain after implantation with the PEEK implant deviated less from the intact case (R2 = 0.7972 with slope ? = 0.939) with no difference (p = 0.231).

CONCLUSIONS: The strain shielding observed with the contemporary CoCr implant, consistent with clinical bone mineral density change data reported by others, may be reduced by using a PEEK implant.

CLINICAL RELEVANCE: This bone analog in vitro study suggests that a PEEK femoral component could transfer more physiologically normal bone strains with a potentially reduced stress shielding effect, which may improve long-term bone preservation. Additional studies including paired cadaver tests are necessary to test the hypothesis further.
0009-921X
2405-2413
Rankin, Kathryn
d9516566-0ad8-473d-b99b-4683c663a2b7
Dickinson, Alexander
10151972-c1b5-4f7d-bc12-6482b5870cad
Briscoe, Adam
40a57aa0-be88-47a6-b528-ae8f0914d0e7
Browne, Martin
6578cc37-7bd6-43b9-ae5c-77ccb7726397
Rankin, Kathryn
d9516566-0ad8-473d-b99b-4683c663a2b7
Dickinson, Alexander
10151972-c1b5-4f7d-bc12-6482b5870cad
Briscoe, Adam
40a57aa0-be88-47a6-b528-ae8f0914d0e7
Browne, Martin
6578cc37-7bd6-43b9-ae5c-77ccb7726397

Rankin, Kathryn, Dickinson, Alexander, Briscoe, Adam and Browne, Martin (2016) Does a PEEK femoral TKA implant preserve intact femoral surface strains compared with CoCr? A preliminary laboratory study. Clinical Orthopaedics and Related Research, 474 (11), 2405-2413. (doi:10.1007/s11999-016-4801-8). (PMID:27020431)

Record type: Article

Abstract

BACKGROUND: Both the material and geometry of a total knee arthroplasty (TKA) component influence the induced periprosthetic bone strain field. Strain, a measure of the local relative deformation in a structure, corresponds to the mechanical stimulus that governs bone remodeling and is therefore a useful in vitro biomechanical measure for assessing the response of bone to new implant designs and materials. A polyetheretherketone (PEEK) femoral implant has the potential to promote bone strains closer to that of natural bone as a result of its low elastic modulus compared with cobalt-chromium (CoCr).

QUESTIONS/PURPOSES: In the present study, we used a Digital Image Correlation (DIC) technique to answer the following question: Does a PEEK TKA femoral component induce a more physiologically normal bone strain distribution than a CoCr component? To achieve this, a DIC test protocol was developed for periprosthetic bone strain assessment using an analog model; the protocol aimed to minimize errors in strain assessment through the selection of appropriate analysis parameters.

METHODS: Three synthetic bone femurs were used in this experiment. One was implanted with a CoCr femoral component and one with a PEEK femoral component. The third (unimplanted) femur was intact and used as the physiological reference (control) model. All models were subjected to standing loads on the corresponding polyethylene (ultrahigh-molecular-weight polyethylene) tibial component, and speckle image data were acquired for surface strain analysis using DIC in six repeat tests. The strain in 16 regions of interest on the lateral surface of each of the implanted bone models was plotted for comparison with the corresponding strains in the intact case. A Wilcoxon signed-rank test was used to test for difference at the 5% significance level.

RESULTS: Surface analog bone strain after CoCr implantation indicated strain shielding (R2 = 0.6178 with slope, ? = 0.4314) and was lower than the intact case (p = 0.014). The strain after implantation with the PEEK implant deviated less from the intact case (R2 = 0.7972 with slope ? = 0.939) with no difference (p = 0.231).

CONCLUSIONS: The strain shielding observed with the contemporary CoCr implant, consistent with clinical bone mineral density change data reported by others, may be reduced by using a PEEK implant.

CLINICAL RELEVANCE: This bone analog in vitro study suggests that a PEEK femoral component could transfer more physiologically normal bone strains with a potentially reduced stress shielding effect, which may improve long-term bone preservation. Additional studies including paired cadaver tests are necessary to test the hypothesis further.

Text
Rankin_CORR_2016 Post Proof.pdf - Accepted Manuscript
Download (1MB)
Text
Rankin_CORR_2016.pdf - Version of Record
Available under License Creative Commons Attribution.
Download (3MB)

More information

Accepted/In Press date: 15 March 2016
e-pub ahead of print date: 28 March 2016
Published date: November 2016
Organisations: Bioengineering Group

Identifiers

Local EPrints ID: 389836
URI: http://eprints.soton.ac.uk/id/eprint/389836
ISSN: 0009-921X
PURE UUID: 8d63c5ea-9cae-47fa-9bf1-d95d08a4db2a
ORCID for Kathryn Rankin: ORCID iD orcid.org/0000-0002-8458-1038
ORCID for Alexander Dickinson: ORCID iD orcid.org/0000-0002-9647-1944
ORCID for Martin Browne: ORCID iD orcid.org/0000-0001-5184-050X

Catalogue record

Date deposited: 07 Apr 2016 15:36
Last modified: 15 Mar 2024 05:26

Export record

Altmetrics

Contributors

Author: Kathryn Rankin ORCID iD
Author: Adam Briscoe
Author: Martin Browne ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×