A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users
A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users
Background
The ankle joint is a common site of musculoskeletal pathology. Measurement of its functional range of motion is a primary indicator for rehabilitation outcomes in therapy settings. The present study was designed to assess reliability and validity of a new standardised method using a D-Flex device to assess ankle range of motion.
Methods
A cohort of 20 healthy volunteers were recruited to measure the weight-bearing ankle range of motion using three assessment tools, namely, a goniometer, inclinometer and the D-Flex measurement devices. Repeated measures were performed both between and within observers for each device over a 48 h period. Performance evaluation of each device and their reliability was assessed using intra-class correlation coefficients and Bland and Altman plots.
Results
Although significant correlations (p < 0.05) were observed between devices, there were large mean differences in ankle range of motion values ranging from 4.3°-15.7°. The D-flex produced the highest inter- and intra-rater reliability (ICCs 0.76–0.95), compared to values of 0.55–0.85 and 0.32–0.71 for the goniometer and inclinometer, respectively. The Bland and Altman plots revealed a low mean observer difference for the D-Flex (mean difference = 0.7°), with the vast majority of data coincident within the 95% confidence intervals. For both the goniometer and inclinometer mean differences were higher, with values of 3.1° and 5.7° respectively.
Conclusion
The results of the present study provide evidence to support the use of the D-Flex system as a valid, portable, and easy to use alternative to the weight-bearing lunge test when assessing ankle dorsiflexion ROM in healthy participants.
1-8
Worsley, Peter
6d33aee3-ef43-468d-aef6-86d190de6756
Conington, Caitlan
0d1aea51-baf2-473b-a0f4-1f6b933010f8
Stuart, Holly
23fcec6d-5c4a-40a4-befd-778ae3003c7b
Patterson, Alice
f050cbb3-dae3-4679-8ec1-15881674fcfe
Bader, Daniel
9884d4f6-2607-4d48-bf0c-62bdcc0d1dbf
Worsley, Peter
6d33aee3-ef43-468d-aef6-86d190de6756
Conington, Caitlan
0d1aea51-baf2-473b-a0f4-1f6b933010f8
Stuart, Holly
23fcec6d-5c4a-40a4-befd-778ae3003c7b
Patterson, Alice
f050cbb3-dae3-4679-8ec1-15881674fcfe
Bader, Daniel
9884d4f6-2607-4d48-bf0c-62bdcc0d1dbf
Worsley, Peter, Conington, Caitlan, Stuart, Holly, Patterson, Alice and Bader, Daniel
(2018)
A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users.
Journal of Foot and Ankle Research, 11 (45), .
(doi:10.1186/s13047-018-0286-x).
Abstract
Background
The ankle joint is a common site of musculoskeletal pathology. Measurement of its functional range of motion is a primary indicator for rehabilitation outcomes in therapy settings. The present study was designed to assess reliability and validity of a new standardised method using a D-Flex device to assess ankle range of motion.
Methods
A cohort of 20 healthy volunteers were recruited to measure the weight-bearing ankle range of motion using three assessment tools, namely, a goniometer, inclinometer and the D-Flex measurement devices. Repeated measures were performed both between and within observers for each device over a 48 h period. Performance evaluation of each device and their reliability was assessed using intra-class correlation coefficients and Bland and Altman plots.
Results
Although significant correlations (p < 0.05) were observed between devices, there were large mean differences in ankle range of motion values ranging from 4.3°-15.7°. The D-flex produced the highest inter- and intra-rater reliability (ICCs 0.76–0.95), compared to values of 0.55–0.85 and 0.32–0.71 for the goniometer and inclinometer, respectively. The Bland and Altman plots revealed a low mean observer difference for the D-Flex (mean difference = 0.7°), with the vast majority of data coincident within the 95% confidence intervals. For both the goniometer and inclinometer mean differences were higher, with values of 3.1° and 5.7° respectively.
Conclusion
The results of the present study provide evidence to support the use of the D-Flex system as a valid, portable, and easy to use alternative to the weight-bearing lunge test when assessing ankle dorsiflexion ROM in healthy participants.
Text
A Randomised Cross Over Study to Evaluate the Performance of a Novel Ankle Dorsiflexion Measurement Device for Novice Users
- Accepted Manuscript
Text
s13047-018-0286-x
- Version of Record
More information
Accepted/In Press date: 19 July 2018
e-pub ahead of print date: 31 July 2018
Identifiers
Local EPrints ID: 423038
URI: http://eprints.soton.ac.uk/id/eprint/423038
ISSN: 1757-1146
PURE UUID: 407ed856-9a77-4275-841c-ba2b37d03366
Catalogue record
Date deposited: 13 Aug 2018 16:30
Last modified: 16 Mar 2024 03:58
Export record
Altmetrics
Contributors
Author:
Caitlan Conington
Author:
Holly Stuart
Author:
Alice Patterson
Download statistics
Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.
View more statistics