A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users

A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users
A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users
Background The ankle joint is a common site of musculoskeletal pathology. Measurement of its functional range of motion is a primary indicator for rehabilitation outcomes in therapy settings. The present study was designed to assess reliability and validity of a new standardised method using a D-Flex device to assess ankle range of motion. Methods A cohort of 20 healthy volunteers were recruited to measure the weight-bearing ankle range of motion using three assessment tools, namely, a goniometer, inclinometer and the D-Flex measurement devices. Repeated measures were performed both between and within observers for each device over a 48 h period. Performance evaluation of each device and their reliability was assessed using intra-class correlation coefficients and Bland and Altman plots. Results Although significant correlations (p < 0.05) were observed between devices, there were large mean differences in ankle range of motion values ranging from 4.3°-15.7°. The D-flex produced the highest inter- and intra-rater reliability (ICCs 0.76–0.95), compared to values of 0.55–0.85 and 0.32–0.71 for the goniometer and inclinometer, respectively. The Bland and Altman plots revealed a low mean observer difference for the D-Flex (mean difference = 0.7°), with the vast majority of data coincident within the 95% confidence intervals. For both the goniometer and inclinometer mean differences were higher, with values of 3.1° and 5.7° respectively. Conclusion The results of the present study provide evidence to support the use of the D-Flex system as a valid, portable, and easy to use alternative to the weight-bearing lunge test when assessing ankle dorsiflexion ROM in healthy participants.
1757-1146
1-8
Worsley, Peter
6d33aee3-ef43-468d-aef6-86d190de6756
Conington, Caitlan
0d1aea51-baf2-473b-a0f4-1f6b933010f8
Stuart, Holly
23fcec6d-5c4a-40a4-befd-778ae3003c7b
Patterson, Alice
f050cbb3-dae3-4679-8ec1-15881674fcfe
Bader, Daniel
9884d4f6-2607-4d48-bf0c-62bdcc0d1dbf
Worsley, Peter
6d33aee3-ef43-468d-aef6-86d190de6756
Conington, Caitlan
0d1aea51-baf2-473b-a0f4-1f6b933010f8
Stuart, Holly
23fcec6d-5c4a-40a4-befd-778ae3003c7b
Patterson, Alice
f050cbb3-dae3-4679-8ec1-15881674fcfe
Bader, Daniel
9884d4f6-2607-4d48-bf0c-62bdcc0d1dbf

Worsley, Peter, Conington, Caitlan, Stuart, Holly, Patterson, Alice and Bader, Daniel (2018) A randomised cross over study to evaluate the performance of a novel ankle dorsiflexion measurement device for novice users. Journal of Foot and Ankle Research, 11 (45), 1-8. (doi:10.1186/s13047-018-0286-x).

Record type: Article

Abstract

Background The ankle joint is a common site of musculoskeletal pathology. Measurement of its functional range of motion is a primary indicator for rehabilitation outcomes in therapy settings. The present study was designed to assess reliability and validity of a new standardised method using a D-Flex device to assess ankle range of motion. Methods A cohort of 20 healthy volunteers were recruited to measure the weight-bearing ankle range of motion using three assessment tools, namely, a goniometer, inclinometer and the D-Flex measurement devices. Repeated measures were performed both between and within observers for each device over a 48 h period. Performance evaluation of each device and their reliability was assessed using intra-class correlation coefficients and Bland and Altman plots. Results Although significant correlations (p < 0.05) were observed between devices, there were large mean differences in ankle range of motion values ranging from 4.3°-15.7°. The D-flex produced the highest inter- and intra-rater reliability (ICCs 0.76–0.95), compared to values of 0.55–0.85 and 0.32–0.71 for the goniometer and inclinometer, respectively. The Bland and Altman plots revealed a low mean observer difference for the D-Flex (mean difference = 0.7°), with the vast majority of data coincident within the 95% confidence intervals. For both the goniometer and inclinometer mean differences were higher, with values of 3.1° and 5.7° respectively. Conclusion The results of the present study provide evidence to support the use of the D-Flex system as a valid, portable, and easy to use alternative to the weight-bearing lunge test when assessing ankle dorsiflexion ROM in healthy participants.

Text
A Randomised Cross Over Study to Evaluate the Performance of a Novel Ankle Dorsiflexion Measurement Device for Novice Users - Accepted Manuscript
Available under License Creative Commons Attribution.
Download (2MB)
Text
s13047-018-0286-x - Version of Record
Available under License Creative Commons Attribution.
Download (911kB)

More information

Accepted/In Press date: 19 July 2018
e-pub ahead of print date: 31 July 2018

Identifiers

Local EPrints ID: 423038
URI: http://eprints.soton.ac.uk/id/eprint/423038
ISSN: 1757-1146
PURE UUID: 407ed856-9a77-4275-841c-ba2b37d03366
ORCID for Peter Worsley: ORCID iD orcid.org/0000-0003-0145-5042
ORCID for Daniel Bader: ORCID iD orcid.org/0000-0002-1208-3507

Catalogue record

Date deposited: 13 Aug 2018 16:30
Last modified: 16 Mar 2024 03:58

Export record

Altmetrics

Contributors

Author: Peter Worsley ORCID iD
Author: Caitlan Conington
Author: Holly Stuart
Author: Alice Patterson
Author: Daniel Bader ORCID iD

Download statistics

Downloads from ePrints over the past year. Other digital versions may also be available to download e.g. from the publisher's website.

View more statistics

Atom RSS 1.0 RSS 2.0

Contact ePrints Soton: eprints@soton.ac.uk

ePrints Soton supports OAI 2.0 with a base URL of http://eprints.soton.ac.uk/cgi/oai2

This repository has been built using EPrints software, developed at the University of Southampton, but available to everyone to use.

We use cookies to ensure that we give you the best experience on our website. If you continue without changing your settings, we will assume that you are happy to receive cookies on the University of Southampton website.

×